Evaluation of ultraviolet C for disinfection of endocavitary ultrasound transducers persistently contaminated despite probe covers
- PMID: 20025531
- DOI: 10.1086/649794
Evaluation of ultraviolet C for disinfection of endocavitary ultrasound transducers persistently contaminated despite probe covers
Abstract
Objective: To determine the rate of bacterial and viral contamination of endocavitary ultrasound probes after endorectal or endovaginal examination with the use of probe covers and to evaluate the antimicrobial efficacy of a disinfection procedure consisting of cleaning with a disinfectant-impregnated towel followed by disinfection with ultraviolet C (UVC) light.
Methods: Endovaginal or endorectal ultrasound examinations were performed for 440 patients in 3 institutions. All probes were covered by a condom or sheath during the examination. For bacterial analysis, 1 swab was applied lengthwise across one-half the surface of the probe just after removal of the probe cover. The second swab was similarly applied over the probe immediately after the end of a 2-step process consisting of cleaning with a towel impregnated with a disinfectant spray and a 5-minute UVC disinfection cycle. Swabs were applied onto plates and incubated for 48 hours. The number of colony-forming units was counted, and organisms were identified. A similar protocol was used for viral detection of Epstein-Barr virus, human cytomegalovirus, and human papillomavirus, except that an additional swab was applied along the entire external surface of the probe cover before its removal. Viruses were detected by means of a polymerase chain reaction-based protocol.
Results: After removal of probe covers, contamination by pathogenic bacteria was found for 15 (3.4% [95% confidence interval, 2.0%-5.6%]) of 440 probes, and viral genome was detected on 5 (1.5% [95% confidence interval, 0.5%-3.5%]) of 336 probes. After cleaning with a towel impregnated with a disinfectant spray and disinfecting with UVC light, neither bacterial pathogenic flora nor viral genome was recovered from the probe.
Conclusions: Endocavitary ultrasound probes may carry pathogens after removal of covers under routine conditions. A disinfection procedure consisting of cleaning with a disinfectant-impregnated towel followed by disinfection with UVC may provide a useful method for disinfecting endocavitary ultrasound probes.
Similar articles
-
Evaluation of a new disinfection procedure for ultrasound probes using ultraviolet light.J Hosp Infect. 2007 Feb;65(2):163-8. doi: 10.1016/j.jhin.2006.10.008. Epub 2006 Dec 14. J Hosp Infect. 2007. PMID: 17174448 Clinical Trial.
-
Evaluation of a new disinfection method for ultrasound probes used for regional anesthesia: ultraviolet C light.J Ultrasound Med. 2011 Jun;30(6):785-8. doi: 10.7863/jum.2011.30.6.785. J Ultrasound Med. 2011. PMID: 21632992
-
Can ultrasound probes and coupling gel be a source of nosocomial infection in patients undergoing sonography? An in vivo and in vitro study.AJR Am J Roentgenol. 1995 Jun;164(6):1521-4. doi: 10.2214/ajr.164.6.7754907. AJR Am J Roentgenol. 1995. PMID: 7754907
-
Ultrasound probes as a possible vector of bacterial transmission.Med Ultrason. 2013 Mar;15(1):41-4. doi: 10.11152/mu.2013.2066.151.hk1upp2. Med Ultrason. 2013. PMID: 23486623 Review.
-
[Hygiene in endovaginal gynecologic and obstetrical ultrasound in 2008].J Gynecol Obstet Biol Reprod (Paris). 2009 Feb;38(1):43-50. doi: 10.1016/j.jgyn.2008.09.009. Epub 2008 Dec 10. J Gynecol Obstet Biol Reprod (Paris). 2009. PMID: 19081685 Review. French.
Cited by
-
Applications of ultraviolet germicidal irradiation disinfection in health care facilities: effective adjunct, but not stand-alone technology.Am J Infect Control. 2010 Jun;38(5 Suppl 1):S13-24. doi: 10.1016/j.ajic.2010.04.208. Am J Infect Control. 2010. PMID: 20569852 Free PMC article. Review.
-
High risk HPV contamination of endocavity vaginal ultrasound probes: an underestimated route of nosocomial infection?PLoS One. 2012;7(10):e48137. doi: 10.1371/journal.pone.0048137. Epub 2012 Oct 24. PLoS One. 2012. PMID: 23110191 Free PMC article.
-
Complete sequencing of pNDM-HK encoding NDM-1 carbapenemase from a multidrug-resistant Escherichia coli strain isolated in Hong Kong.PLoS One. 2011 Mar 21;6(3):e17989. doi: 10.1371/journal.pone.0017989. PLoS One. 2011. PMID: 21445317 Free PMC article.
-
Ultrasound scan as a potential source of nosocomial and crossinfection: a literature review.Radiol Bras. 2015 Sep-Oct;48(5):319-23. doi: 10.1590/0100-3984.2014.0002. Radiol Bras. 2015. PMID: 26543284 Free PMC article. Review.
-
Decontamination of targeted pathogens from patient rooms using an automated ultraviolet-C-emitting device.Infect Control Hosp Epidemiol. 2013 May;34(5):466-71. doi: 10.1086/670215. Infect Control Hosp Epidemiol. 2013. PMID: 23571362 Free PMC article.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Medical