Effect of monitor display on detection of approximal caries lesions in digital radiographs
- PMID: 20026711
- DOI: 10.1259/dmfr/21071028
Effect of monitor display on detection of approximal caries lesions in digital radiographs
Abstract
Objectives: The aim was to compare the accuracy of five flat panel monitors for detection of approximal caries lesions.
Methods: Five flat panel monitors, Mermaid Ventura (15 inch, colour flat panel, 1024 x 768, 32 bit, analogue), Olórin VistaLine (19 inch, colour, 1280 x 1024, 32 bit, digital), Samsung SyncMaster 203B (20 inch, colour, 1024 x 768, 32 bit, analogue), Totoku ME251i (21 inch, greyscale, 1400 x 1024, 32 bit, digital) and Eizo FlexScan MX190 (19 inch, colour, 1280 x 1024, 32 bit, digital), were assessed. 160 approximal surfaces of human teeth were examined with a storage phosphor plate system (Digora FMX, Soredex) and assessed by seven observers for the presence of caries lesions. Microscopy of the teeth served as validation for the presence/absence of a lesion.
Results: The sensitivities varied between observers (range 7-25%) but the variation between the monitors was not large. The Samsung monitor obtained a significantly higher sensitivity than the Mermaid and Olórin monitors (P<0.02) and a lower specificity than the Eizo and Totoku monitors (P<0.05). There were no significant differences between any other monitors. The percentage of correct scores was highest for the Eizo monitor and significantly higher than for the Mermaid and Olórin monitors (P<0.03).
Conclusions: There was no clear relationship between the diagnostic accuracy and the resolution or price of the monitor. The Eizo monitor was associated with the overall highest percentage of correct scores. The standard analogue flat panel monitor, Samsung, had higher sensitivity and lower specificity than some of the other monitors, but did not differ in overall accuracy for detection of carious lesions.
Similar articles
-
Diagnostic accuracy of two software modalities for detection of caries lesions in digital radiographs from four dental systems.Dentomaxillofac Radiol. 2006 Mar;35(2):78-82. doi: 10.1259/dmfr/50356588. Dentomaxillofac Radiol. 2006. PMID: 16549433
-
Effect of ambient light level at the monitor surface on digital radiographic evaluation of approximal carious lesions: an in vitro study.Dentomaxillofac Radiol. 2012 Mar;41(3):192-6. doi: 10.1259/dmfr/15422221. Dentomaxillofac Radiol. 2012. PMID: 22378753 Free PMC article.
-
Accuracy of caries detection with four storage phosphor systems and E-speed radiographs.Dentomaxillofac Radiol. 2002 May;31(3):170-5. doi: 10.1038/sj/dmfr/4600686. Dentomaxillofac Radiol. 2002. PMID: 12058264
-
Detection of artificial occlusal caries in a phosphor imaging plate system with two types of LCD monitors versus three different films.J Digit Imaging. 2009 Jun;22(3):242-9. doi: 10.1007/s10278-008-9146-8. Epub 2008 Oct 24. J Digit Imaging. 2009. PMID: 18949518 Free PMC article.
-
Diagnostic accuracy of cone beam computed tomography scans compared with intraoral image modalities for detection of caries lesions.Dentomaxillofac Radiol. 2008 Jan;37(1):18-22. doi: 10.1259/dmfr/87103878. Dentomaxillofac Radiol. 2008. PMID: 18195250
Cited by
-
Imaging modalities to inform the detection and diagnosis of early caries.Cochrane Database Syst Rev. 2021 Mar 15;3(3):CD014545. doi: 10.1002/14651858.CD014545. Cochrane Database Syst Rev. 2021. PMID: 33720395 Free PMC article.
-
Influence of materials radiopacity in the radiographic diagnosis of secondary caries: evaluation in film and two digital systems.Dentomaxillofac Radiol. 2011 Sep;40(6):344-50. doi: 10.1259/dmfr/93764866. Dentomaxillofac Radiol. 2011. PMID: 21831973 Free PMC article.
-
Caries detection under composite restorations by laser fluorescence and digital radiography.Clin Oral Investig. 2013 Dec;17(9):2079-84. doi: 10.1007/s00784-012-0908-9. Epub 2012 Dec 16. Clin Oral Investig. 2013. PMID: 23242816
-
Measurements of simulated periodontal bone defects in inverted digital image and film-based radiograph: an in vitro study.Imaging Sci Dent. 2012 Dec;42(4):243-7. doi: 10.5624/isd.2012.42.4.243. Epub 2012 Dec 23. Imaging Sci Dent. 2012. PMID: 23301211 Free PMC article.
-
Comparison between DICOM-calibrated and uncalibrated consumer grade and 6-MP displays under different lighting conditions in panoramic radiography.Dentomaxillofac Radiol. 2015;44(5):20140365. doi: 10.1259/dmfr.20140365. Epub 2015 Jan 7. Dentomaxillofac Radiol. 2015. PMID: 25564888 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical