B1 transmission-field inhomogeneity and enhancement ratio errors in dynamic contrast-enhanced MRI (DCE-MRI) of the breast at 3T
- PMID: 20027594
- DOI: 10.1002/jmri.22018
B1 transmission-field inhomogeneity and enhancement ratio errors in dynamic contrast-enhanced MRI (DCE-MRI) of the breast at 3T
Abstract
Purpose: To quantify B(1) transmission-field inhomogeneity in breast imaging of normal volunteers at 3T using 3D T(1)-weighted spoiled gradient echo and to assess the resulting errors in enhancement ratio (ER) measured in dynamic contrast-enhanced MRI (DCE-MRI) studies of the breast.
Materials and methods: A total of 25 volunteers underwent breast imaging at 3T and the B(1) transmission-fields were mapped. Gel phantoms that simulate pre- and postcontrast breast tissue T(1) were developed. The effects of B(1)-field inhomogeneity on ER, as measured using a 3D spoiled gradient echo sequence, were investigated by computer simulation and experiments on gel phantoms.
Results: It was observed that by using the patient orientation and MR scanner employed in this study, the B(1) transmission-field field is always reduced toward the volunteer's right side. The median B(1)-field in the right breast is reduced around 40% of the expected B(1)-field. For some volunteers the amplitude was reduced by more than 50%. Computer simulation and experiment showed that a reduction in B(1)-field decreases ER. This reduction increases with both B(1)-field error and contrast agent uptake.
Conclusion: B(1) transmission-field inhomogeneity is a critical issue in breast imaging at 3T and causes errors in quantifying ER. These errors would be sufficient to reduce the conspicuity of a malignant lesion and could result in reduced sensitivity for cancer detection.
(c) 2009 Wiley-Liss, Inc.
Similar articles
-
Transmit B1+ field inhomogeneity and T1 estimation errors in breast DCE-MRI at 3 tesla.J Magn Reson Imaging. 2013 Aug;38(2):454-9. doi: 10.1002/jmri.23996. Epub 2013 Jan 4. J Magn Reson Imaging. 2013. PMID: 23292822 Free PMC article.
-
Characterizing and eliminating errors in enhancement and subtraction artifacts in dynamic contrast-enhanced breast MRI: Chemical shift artifact of the third kind.Magn Reson Med. 2018 Apr;79(4):2277-2289. doi: 10.1002/mrm.26879. Epub 2017 Aug 24. Magn Reson Med. 2018. PMID: 28840613 Free PMC article.
-
Practical methods for improving B1+ homogeneity in 3 Tesla breast imaging.J Magn Reson Imaging. 2015 Apr;41(4):992-9. doi: 10.1002/jmri.24635. Epub 2014 Apr 10. J Magn Reson Imaging. 2015. PMID: 24723508 Free PMC article.
-
Automatic identification and classification of characteristic kinetic curves of breast lesions on DCE-MRI.Med Phys. 2006 Aug;33(8):2878-87. doi: 10.1118/1.2210568. Med Phys. 2006. PMID: 16964864
-
Comparison of conventional DCE-MRI and a novel golden-angle radial multicoil compressed sensing method for the evaluation of breast lesion conspicuity.J Magn Reson Imaging. 2017 Jun;45(6):1746-1752. doi: 10.1002/jmri.25530. Epub 2016 Nov 17. J Magn Reson Imaging. 2017. PMID: 27859874 Free PMC article. Clinical Trial.
Cited by
-
Effects of flip angle uncertainty and noise on the accuracy of DCE-MRI metrics: comparison between standard concentration-based and signal difference methods.Magn Reson Imaging. 2015 Jan;33(1):166-73. doi: 10.1016/j.mri.2014.10.005. Epub 2014 Oct 13. Magn Reson Imaging. 2015. PMID: 25311569 Free PMC article.
-
Evaluation of B1 inhomogeneity effect on DCE-MRI data analysis of brain tumor patients at 3T.J Transl Med. 2017 Dec 2;15(1):242. doi: 10.1186/s12967-017-1349-7. J Transl Med. 2017. PMID: 29197390 Free PMC article.
-
Improved B1 homogeneity of 3 Tesla breast MRI using dual-source parallel radiofrequency excitation.J Magn Reson Imaging. 2012 May;35(5):1222-6. doi: 10.1002/jmri.23571. Epub 2012 Jan 26. J Magn Reson Imaging. 2012. PMID: 22282269 Free PMC article.
-
Measurement of murine kidney functional biomarkers using DCE-MRI: A multi-slice TRICKS technique and semi-automated image processing algorithm.Magn Reson Imaging. 2019 Nov;63:226-234. doi: 10.1016/j.mri.2019.08.029. Epub 2019 Aug 20. Magn Reson Imaging. 2019. PMID: 31442558 Free PMC article.
-
Radiofrequency transmit calibration: A multi-center evaluation of vendor-provided radiofrequency transmit mapping methods.Med Phys. 2019 Jun;46(6):2629-2637. doi: 10.1002/mp.13518. Epub 2019 Apr 15. Med Phys. 2019. PMID: 30924940 Free PMC article.
Publication types
MeSH terms
Substances
Grants and funding
LinkOut - more resources
Full Text Sources
Medical