B1 transmission-field inhomogeneity and enhancement ratio errors in dynamic contrast-enhanced MRI (DCE-MRI) of the breast at 3T

Affiliations

B1 transmission-field inhomogeneity and enhancement ratio errors in dynamic contrast-enhanced MRI (DCE-MRI) of the breast at 3T

Che A Azlan et al. J Magn Reson Imaging. 2010 Jan.

Abstract

Purpose: To quantify B(1) transmission-field inhomogeneity in breast imaging of normal volunteers at 3T using 3D T(1)-weighted spoiled gradient echo and to assess the resulting errors in enhancement ratio (ER) measured in dynamic contrast-enhanced MRI (DCE-MRI) studies of the breast.

Materials and methods: A total of 25 volunteers underwent breast imaging at 3T and the B(1) transmission-fields were mapped. Gel phantoms that simulate pre- and postcontrast breast tissue T(1) were developed. The effects of B(1)-field inhomogeneity on ER, as measured using a 3D spoiled gradient echo sequence, were investigated by computer simulation and experiments on gel phantoms.

Results: It was observed that by using the patient orientation and MR scanner employed in this study, the B(1) transmission-field field is always reduced toward the volunteer's right side. The median B(1)-field in the right breast is reduced around 40% of the expected B(1)-field. For some volunteers the amplitude was reduced by more than 50%. Computer simulation and experiment showed that a reduction in B(1)-field decreases ER. This reduction increases with both B(1)-field error and contrast agent uptake.

Conclusion: B(1) transmission-field inhomogeneity is a critical issue in breast imaging at 3T and causes errors in quantifying ER. These errors would be sufficient to reduce the conspicuity of a malignant lesion and could result in reduced sensitivity for cancer detection.