Comparative Study
. 2010 Jan;194(1):W60-4.
doi: 10.2214/AJR.08.1740.

Imaging of soft tissues adjacent to orthopedic hardware: comparison of 3-T and 1.5-T MRI

Affiliations
Comparative Study

Imaging of soft tissues adjacent to orthopedic hardware: comparison of 3-T and 1.5-T MRI

Cormac Farrelly et al. AJR Am J Roentgenol. 2010 Jan.

Erratum in

  • AJR Am J Roentgenol. 2012 Aug;199(2):471. Brennan, Stephen [corrected to Brennan, Stephen A]

Abstract

Objective: The purpose of this study was to compare metal artifact reduction techniques at 1.5-T and 3-T MRI.

Materials and methods: A titanium plate with steel screws was placed in a freshly harvested pig leg. The leg was imaged with 1.5-T and 3-T MRI. A T2-weighted turbo spin-echo sequence was used with echo-train lengths of 8, 16, 32, and 64 and a constant readout bandwidth of 31.2 kHz. The images were compared qualitatively, and the optimal echo-train length was selected. Images were acquired at the optimal echo-train length with four different readout bandwidths. Artifact was measured quantitatively, and image quality was ranked qualitatively. The qualitatively best image acquired at 1.5 T was compared with the qualitatively highest-ranked image acquired at 3 T.

Results: At both 1.5 T and 3 T, optimal images of equal quality were produced at echo-train lengths of 8 and 16. At higher readout bandwidths, there was quantitatively less artifact. The qualitatively best images were acquired at a readout bandwidth of 31.2 kHz at 1.5 T and 62.5 kHz at 3 T (Cronbach's alpha=1.00). The optimal image at 3 T was qualitatively superior to that at 1.5 T.

Conclusion: Optimizing image acquisition parameters in this phantom model resulted in similar quantitative susceptibility artifact at 3 T and 1.5 T and better qualitative images at 3 T than at 1.5 T.