Dual-energy CT angiography in the evaluation of intracranial aneurysms: image quality, radiation dose, and comparison with 3D rotational digital subtraction angiography
- PMID: 20028901
- DOI: 10.2214/AJR.08.2290
Dual-energy CT angiography in the evaluation of intracranial aneurysms: image quality, radiation dose, and comparison with 3D rotational digital subtraction angiography
Abstract
Objective: The purpose of this study was to evaluate the image quality, radiation dose, and diagnostic accuracy of dual-energy CT angiography (CTA) compared with 3D rotational digital subtraction angiography (DSA) in the detection of intracranial aneurysms.
Subjects and methods: Forty-six patients with clinically suspected intracranial aneurysms underwent dual-source dual-energy CTA and 3D DSA. For the analysis of the image quality and radiation dose of dual-energy CTA, 46 patients who underwent digital subtraction CTA were recruited as a control group. The image quality of dual-energy CTA and digital subtraction CTA was rated on a 4-point scale as excellent, good, moderate, or poor. The radiation dose of CTA was recorded according to patient protocol. Aneurysm detection with dual-energy CTA compared with 3D DSA was analyzed on a per-patient and on a peraneurysm basis. Sensitivity, specificity, and positive and negative predictive values for aneurysm presence were determined. The mean maximum diameter and dome and neck dimensions of aneurysms were measured on dual-energy CTA and 3D DSA images. Correlation analysis between the two techniques was performed.
Results: There was no statistical difference between the image quality of dual-energy CTA and that of digital subtraction CTA (p>0.05). Patients undergoing dual-energy CTA received a smaller radiation dose (volume CT dose index, 20.6+/-0.1 mGy [mean+/-SD]; dose-length product, 398.6+/-19.0 mGy x cm) than those undergoing digital subtraction CTA (volume CT dose index, 50.4+/-3.4 mGy; dose-length product, 1,095.6+/-114.2 mGyxcm) (p<0.05). Three-dimensional DSA showed no aneurysm in 11 patients and 40 aneurysms in 35 patients. The mean maximum diameter of the aneurysms was 6+/-3 mm; the dome measurement, 5+/-3 mm; and the neck dimension, 3+/-2 mm. With dual-energy CTA, 38 aneurysms in 34 patients were correctly detected, and two aneurysms in two patients were missed. With DSA as the standard of reference, the sensitivity, specificity, and positive and negative predictive values of dual-energy CTA in the detection of intracranial aneurysm were 97.1%, 100%, 100%, and 91.7% on a per-patient basis and 95.0%, 100%, 100%, and 99.7% on a per-aneurysm basis. Dual-energy CTA had sensitivities of 93.8%, 100%, and 80.0% and specificities of 100%, 100%, and 100% in the detection of aneurysms larger than 5 mm, those measuring 3.1-5 mm, and aneurysms 3 mm or smaller. At dual-energy CTA, the mean maximum diameter and dome and neck dimensions were 6+/-3 mm, 5+/-3 mm, and 3+/-2 mm. Excellent correlation was found between DSA and dual-energy CTA findings with respect to mean maximum diameter and dome and neck dimensions (r=0.969, 0.957, and 0.870; p = 0.000).
Conclusion: On the basis of the findings in the small series of patients evaluated, contrast-enhanced dual-energy CTA had diagnostic image quality at a lower radiation dose than digital subtraction CTA and high diagnostic accuracy compared with 3D DSA in the detection of intracranial aneurysms.
Similar articles
-
Digital subtraction CT angiography for detection of intracranial aneurysms: comparison with three-dimensional digital subtraction angiography.Radiology. 2012 Feb;262(2):605-12. doi: 10.1148/radiol.11110486. Epub 2011 Dec 5. Radiology. 2012. PMID: 22143927
-
3D bone subtraction CT angiography for the evaluation of intracranial aneurysms: a comparison study with 2D bone subtraction CT angiography and conventional non-subtracted CT angiography.Acta Radiol. 2015 Sep;56(9):1127-34. doi: 10.1177/0284185114549495. Epub 2014 Sep 25. Acta Radiol. 2015. PMID: 25258207
-
Automatic bone removal dual-energy CT angiography for the evaluation of intracranial aneurysms.J Comput Assist Tomogr. 2010 Nov-Dec;34(6):816-24. doi: 10.1097/RCT.0b013e3181eff93c. J Comput Assist Tomogr. 2010. PMID: 21084894
-
An update on three-dimensional ct angiography in aneurysms: a useful modality for a neurosurgeon.Turk Neurosurg. 2013;23(3):304-11. doi: 10.5137/1019-5149.JTN.5958-12.0. Turk Neurosurg. 2013. PMID: 23756967 Review.
-
Can computed tomography angiography of the brain replace lumbar puncture in the evaluation of acute-onset headache after a negative noncontrast cranial computed tomography scan?Acad Emerg Med. 2010 Apr;17(4):444-51. doi: 10.1111/j.1553-2712.2010.00694.x. Acad Emerg Med. 2010. PMID: 20370785 Review.
Cited by
-
Dual-Energy CT: What the Neuroradiologist Should Know.Curr Radiol Rep. 2015;3(5):16. doi: 10.1007/s40134-015-0097-9. Curr Radiol Rep. 2015. PMID: 25815242 Free PMC article. Review.
-
Comparison of lower extremity atherosclerosis in diabetic and non-diabetic patients using multidetector computed tomography.BMC Cardiovasc Disord. 2014 Sep 24;14:125. doi: 10.1186/1471-2261-14-125. BMC Cardiovasc Disord. 2014. PMID: 25252783 Free PMC article.
-
Long-term follow-up survey reveals a high yield, up to 30% of patients presenting newly detected aneurysms more than 10 years after ruptured intracranial aneurysms clipping.Neurosurg Rev. 2011 Oct;34(4):485-96. doi: 10.1007/s10143-011-0332-7. Epub 2011 Jun 4. Neurosurg Rev. 2011. PMID: 21643681
-
Applications of spectral CT in emergency department imaging, with a focus on neuroimaging.Emerg Radiol. 2025 Aug;32(4):591-604. doi: 10.1007/s10140-025-02358-x. Epub 2025 Jun 14. Emerg Radiol. 2025. PMID: 40514585 Review.
-
Single Phase Dual-energy CT Angiography: One-stop-shop Tool for Evaluating Aneurysmal Subarachnoid Hemorrhage.Sci Rep. 2016 May 25;6:26704. doi: 10.1038/srep26704. Sci Rep. 2016. PMID: 27222163 Free PMC article. Clinical Trial.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Medical