Automated insertion of preformed cochlear implant electrodes: evaluation of curling behaviour and insertion forces on an artificial cochlear model
- PMID: 20033522
- DOI: 10.1007/s11548-009-0299-9
Automated insertion of preformed cochlear implant electrodes: evaluation of curling behaviour and insertion forces on an artificial cochlear model
Abstract
Purpose: As a substantial part of our concept of a minimally invasive cochlear implant (CI) surgery, we developed an automated insertion tool. Studies on an artificial scala tympani model were performed in order to evaluate force application when using the insertion tool.
Methods: Contour electrodes were automatically inserted into a transparent cochlea model in Advance Off-Stylet technique. Occurring forces were measured by the use of a load cell and correlated with observed intracochlear movement of the electrode carriers.
Results: Mean insertion forces were measured up to 20 mN comparable to previous studies on temporal bones. The most influencing factor is the implant's 2D curling behaviour in comparison to the 3D helical shape of the cochlea.
Conclusion: The study confirms the functionality and reliability of the automated insertion tool for insertion of preformed CI. Improved insertion strategies considering patient-specific anatomy become possible.
Similar articles
-
Determination of the curling behavior of a preformed cochlear implant electrode array.Int J Comput Assist Radiol Surg. 2011 May;6(3):421-33. doi: 10.1007/s11548-010-0520-x. Epub 2010 Jul 28. Int J Comput Assist Radiol Surg. 2011. PMID: 20665247
-
Force of cochlear implant electrode insertion performed by a robotic insertion tool: comparison of traditional versus Advance Off-Stylet techniques.Otol Neurotol. 2010 Oct;31(8):1207-10. doi: 10.1097/MAO.0b013e3181f2ebc3. Otol Neurotol. 2010. PMID: 20814345 Free PMC article.
-
An automated insertion tool for cochlear implants: another step towards atraumatic cochlear implant surgery.Int J Comput Assist Radiol Surg. 2010 Mar;5(2):163-71. doi: 10.1007/s11548-009-0368-0. Epub 2009 Jun 13. Int J Comput Assist Radiol Surg. 2010. PMID: 20033518
-
An overview of cochlear implant electrode array designs.Hear Res. 2017 Dec;356:93-103. doi: 10.1016/j.heares.2017.10.005. Epub 2017 Oct 18. Hear Res. 2017. PMID: 29102129 Review.
-
Hyrtl's fissures in cochlear implant surgery.Cochlear Implants Int. 2021 Mar;22(2):116-120. doi: 10.1080/14670100.2020.1823127. Epub 2020 Sep 23. Cochlear Implants Int. 2021. PMID: 32967593 Review.
Cited by
-
A PLLA Coating Does Not Affect the Insertion Pressure or Frictional Behavior of a CI Electrode Array at Higher Insertion Speeds.Materials (Basel). 2022 Apr 22;15(9):3049. doi: 10.3390/ma15093049. Materials (Basel). 2022. PMID: 35591381 Free PMC article.
-
Individual Optimization of the Insertion of a Preformed Cochlear Implant Electrode Array.Int J Otolaryngol. 2015;2015:724703. doi: 10.1155/2015/724703. Epub 2015 Sep 10. Int J Otolaryngol. 2015. PMID: 26448764 Free PMC article.
-
Cochlear Dummy Electrodes for Insertion Training and Research Purposes: Fabrication, Mechanical Characterization, and Experimental Validation.Biomed Res Int. 2015;2015:574209. doi: 10.1155/2015/574209. Epub 2015 Jul 5. Biomed Res Int. 2015. PMID: 26247024 Free PMC article.
-
Intracochlear Pressure Transients During Cochlear Implant Electrode Insertion: Effect of Micro-mechanical Control on Limiting Pressure Trauma.Otol Neurotol. 2019 Jul;40(6):736-744. doi: 10.1097/MAO.0000000000002164. Otol Neurotol. 2019. PMID: 31192901 Free PMC article.
-
Patient-specific estimation of detailed cochlear shape from clinical CT images.Int J Comput Assist Radiol Surg. 2018 Mar;13(3):389-396. doi: 10.1007/s11548-017-1701-7. Epub 2018 Jan 6. Int J Comput Assist Radiol Surg. 2018. PMID: 29305790
References
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Research Materials