Accuracy of stroke volume variation compared with pleth variability index to predict fluid responsiveness in mechanically ventilated patients undergoing major surgery
- PMID: 20035228
- DOI: 10.1097/EJA.0b013e328335fbd1
Accuracy of stroke volume variation compared with pleth variability index to predict fluid responsiveness in mechanically ventilated patients undergoing major surgery
Abstract
Background and objective: Accurate assessment of a patient's volume status is an important goal for an anaesthetist. However, most variables assessing fluid responsiveness are either invasive or technically challenging. This study was designed to compare the accuracy of arterial pressure-based stroke volume variation (SVV) and variations in the pulse oximeter plethysmographic waveform amplitude as evaluated with the noninvasive calculated pleth variability index (PVI) with central venous pressure to predict the response of stroke volume index (SVI) to volume replacement in patients undergoing major surgery.
Methods: We studied 20 patients scheduled for elective major abdominal surgery. After induction of anaesthesia, all haemodynamic variables were recorded immediately before (T1) and subsequent to volume replacement (T2) by infusion of 6% hydroxy-ethyl starch (HES) 130/0.4 (7 ml kg) at a rate of 1 ml kg min.
Results: The volume-induced increase in SVI was at least 15% in 15 patients (responders) and less than 15% in five patients (nonresponders). Baseline SVV correlated significantly with changes in SVI (DeltaSVI; r = 0.80; P < 0.001) as did baseline PVI (r = 0.61; P < 0.004), whereas baseline values of central venous pressure showed no correlation to DeltaSVI. There was no significant difference between the area under the receiver operating characteristic curve for SVV (0.993) and PVI (0.973). The best threshold values to predict fluid responsiveness were more than 11% for SVV and more than 9.5% for PVI.
Conclusion: Although arterial pressure-derived SVV revealed the best correlation to volume-induced changes in SVI, the results of our study suggest that both variables, SVV and PVI, can serve as valid indicators of fluid responsiveness in mechanically ventilated patients undergoing major surgery.
Similar articles
-
Accuracy of the pleth variability index to predict fluid responsiveness depends on the perfusion index.Acta Anaesthesiol Scand. 2011 Jul;55(6):686-93. doi: 10.1111/j.1399-6576.2011.02435.x. Epub 2011 Apr 11. Acta Anaesthesiol Scand. 2011. PMID: 21480831
-
Dynamic variables of fluid responsiveness during pneumoperitoneum and laparoscopic surgery.Acta Anaesthesiol Scand. 2012 Jul;56(6):777-86. doi: 10.1111/j.1399-6576.2011.02641.x. Epub 2012 Jan 31. Acta Anaesthesiol Scand. 2012. PMID: 22288953
-
The ability of pulse pressure variations obtained with CNAP™ device to predict fluid responsiveness in the operating room.Anesth Analg. 2011 Sep;113(3):523-8. doi: 10.1213/ANE.0b013e3182240054. Epub 2011 Jun 3. Anesth Analg. 2011. PMID: 21642606
-
Using ventilation-induced plethysmographic variations to optimize patient fluid status.Curr Opin Anaesthesiol. 2008 Dec;21(6):772-8. doi: 10.1097/ACO.0b013e32831504ca. Curr Opin Anaesthesiol. 2008. PMID: 18997528 Review.
-
Assessment of fluid responsiveness using pulse pressure variation, stroke volume variation, plethysmographic variability index, central venous pressure, and inferior vena cava variation in patients undergoing mechanical ventilation: a systematic review and meta-analysis.Crit Care. 2024 Aug 31;28(1):289. doi: 10.1186/s13054-024-05078-9. Crit Care. 2024. PMID: 39217370 Free PMC article.
Cited by
-
Peripheral Perfusion Index for Prediction of Fluid Responsiveness in Spontaneously Breathing Critically Ill Patients: A Prospective Observational Study.Indian J Crit Care Med. 2025 Feb;29(2):151-154. doi: 10.5005/jp-journals-10071-24898. Epub 2025 Jan 31. Indian J Crit Care Med. 2025. PMID: 40110150 Free PMC article.
-
Accuracy of plethysmographic indices as predictors of fluid responsiveness in mechanically ventilated adults: a systematic review and meta-analysis.Intensive Care Med. 2012 Sep;38(9):1429-37. doi: 10.1007/s00134-012-2621-1. Epub 2012 Jun 26. Intensive Care Med. 2012. PMID: 22732902
-
Prediction of volume responsiveness using pleth variability index in patients undergoing cardiac surgery after cardiopulmonary bypass.J Anesth. 2012 Oct;26(5):696-701. doi: 10.1007/s00540-012-1410-x. Epub 2012 May 16. J Anesth. 2012. PMID: 22588287
-
Evaluation of cardiac function using heart-lung interactions.Ann Transl Med. 2018 Sep;6(18):356. doi: 10.21037/atm.2018.08.10. Ann Transl Med. 2018. PMID: 30370283 Free PMC article. Review.
-
Reliability of pleth variability index in predicting preload responsiveness of mechanically ventilated patients under various conditions: a systematic review and meta-analysis.BMC Anesthesiol. 2019 May 8;19(1):67. doi: 10.1186/s12871-019-0744-4. BMC Anesthesiol. 2019. PMID: 31068139 Free PMC article.