Abstract
Objective: The current standard of manual chest compression during cardiopulmonary resuscitation requires pauses for rhythm analysis and shock delivery. However, interruptions of chest compression greatly decrease the likelihood of successful defibrillations, and significantly better outcomes are reported if this interruption is avoided. We therefore undertook a prospective randomized controlled animal study in an electrically induced ventricular fibrillation pig model to assess the effects of timing of defibrillation on the manual chest compression cycle on the defibrillation threshold.
Design: Prospective, randomized, controlled animal study.
Setting: University-affiliated research laboratory.
Subjects: Yorkshire-X domestic pigs (Sus scrofa).
Interventions: In eight domestic male pigs weighing between 24 and 31 kg, ventricular fibrillation was electrically induced and untreated for 10 secs. Manual chest compression was then performed and continued for 25 secs with the protection of an isolation blanket. The depth and frequency of chest compressions were guided by a cardiopulmonary resuscitation prompter. Animals were randomized to receive a biphasic electrical shock in five different compression phases with a predetermined energy setting. A control phase was chosen at a constant 2 secs after discontinued chest compression. A grouped up-down defibrillation threshold testing protocol was used to compare the success rate at different coupling phases. After a recovery interval of 4 mins, the sequence was repeated for a total of 60 test shocks for each animal.
Measurements and main results: No difference in coronary perfusion pressure before delivering of the shock was observed among the six study phases. The defibrillation success rate, however, was significantly higher when shocks were delivered in the upstroke phase of manual chest compression.
Conclusion: Defibrillation efficacy is maximal when electrical shock is delivered during the upstroke phase of manual chest compression.
Comment in
-
The ups and downs of cardiac defibrillation: integrating the shock with cardiopulmonary resuscitation for success.Crit Care Med. 2010 Mar;38(3):1005-6. doi: 10.1097/CCM.0b013e3181d16b00. Crit Care Med. 2010. PMID: 20168169 No abstract available.
-
The effect of lung volume on efficacy of hands-on defibrillation during each phase of ongoing manual cardiopulmonary resuscitation.Crit Care Med. 2010 May;38(5):1392-3; author reply 1393. doi: 10.1097/CCM.0b013e3181d53a1d. Crit Care Med. 2010. PMID: 20404644 No abstract available.
Similar articles
-
The optimal phasic relationship between synchronized shock and mechanical chest compressions.Resuscitation. 2010 Jun;81(6):724-9. doi: 10.1016/j.resuscitation.2010.02.022. Epub 2010 Mar 25. Resuscitation. 2010. PMID: 20346567
-
Minimal interruption of cardiopulmonary resuscitation for a single shock as mandated by automated external defibrillations does not compromise outcomes in a porcine model of cardiac arrest and resuscitation.Crit Care Med. 2008 Nov;36(11):3048-53. doi: 10.1097/CCM.0b013e318186f612. Crit Care Med. 2008. PMID: 18824916
-
Initial defibrillation versus initial chest compression in a 4-minute ventricular fibrillation canine model of cardiac arrest.Crit Care Med. 2009 Jul;37(7):2250-2. doi: 10.1097/CCM.0b013e31819ffc6a. Crit Care Med. 2009. PMID: 19455026
-
Defibrillation in clinical practice.Curr Opin Crit Care. 2009 Jun;15(3):209-15. doi: 10.1097/MCC.0b013e32832931cb. Curr Opin Crit Care. 2009. PMID: 19262373 Review.
-