A novel noninvasive ultrasonic cardiac output monitor: comparison with cardiac magnetic resonance
- PMID: 20043339
- PMCID: PMC6653547
- DOI: 10.1002/clc.20612
A novel noninvasive ultrasonic cardiac output monitor: comparison with cardiac magnetic resonance
Abstract
Background: USCOM, a novel continuous wave Doppler (CWD) device, has been introduced for noninvasive determination of cardiac output (CO). The present study aimed to compare the accuracy and reproducibility of the new device, using cardiovascular magnetic resonance imaging (CMR) as the noninvasive gold standard.
Methods and results: The CO of 56 consecutive patients was prospectively determined by CWD either before or after CMR imaging. The CWD probe was placed in the suprasternal or supraclavicular notch aiming at the aortic valve. Valid CWD signals could be obtained in 45 patients yielding a CO of 5.3+/-1.1 L/min (range, 3.0-7.5 L/min) by CMR and 4.7+/-1.1 L/min by CWD (2.5-8.0 L/min, P = .004), respectively. CWD measurements showed an acceptable agreement with CMR (bias: 0.6+/-1.1 L/min) and a high reproducibility (bias: 0.1+/-0.4 L/min). Higher CO and body mass index (BMI) were identified as sources of inaccuracy in univariate analysis. By multivariate analysis, only CO(CMR) was found to be independently associated with larger variation. Estimated diameters of the left ventricular outflow tract (LVOT), a prerequisite for CO measurement by CWD, correlated only weakly with those measured by CMR.
Conclusions: Continuous wave Doppler is a feasible technique for measuring cardiac function. Although the overall agreement with CMR was acceptable, CWD showed a trend to underestimate CO. The estimated LVOT diameter by CWD is likely to be an important source of error. Nevertheless, the CWD device could be of clinical use especially for detection of intraindividual hemodynamic changes since a high reproducibility could be demonstrated.
Copyright 2010 Wiley Periodicals, Inc.
Similar articles
-
[Clinical study of neonatal cardiac output measurement methods].Zhonghua Er Ke Za Zhi. 2013 Jan;51(1):58-63. Zhonghua Er Ke Za Zhi. 2013. PMID: 23527933 Chinese.
-
Comparison of continous-wave Doppler ultrasound monitor and echocardiography to assess cardiac output in intensive care patients.Crit Care Resusc. 2017 Sep;19(3):222-229. Crit Care Resusc. 2017. PMID: 28866972
-
Noninvasive cardiac output monitoring by aortic blood flow determination: evaluation of the Sometec Dynemo-3000 system.Crit Care Med. 1998 Dec;26(12):2066-72. doi: 10.1097/00003246-199812000-00043. Crit Care Med. 1998. PMID: 9875922
-
Is cardiac magnetic resonance imaging as accurate as echocardiography in the assessment of aortic valve stenosis?Interact Cardiovasc Thorac Surg. 2016 Apr;22(4):480-6. doi: 10.1093/icvts/ivv362. Epub 2015 Dec 30. Interact Cardiovasc Thorac Surg. 2016. PMID: 26718319 Review.
-
A meta-analysis of the accuracy and precision of the ultrasonic cardiac output monitor (USCOM).Anaesthesia. 2012 Nov;67(11):1266-71. doi: 10.1111/j.1365-2044.2012.07311.x. Epub 2012 Aug 29. Anaesthesia. 2012. PMID: 22928650 Review.
Cited by
-
Comparison of electrical velocimetry and cardiac magnetic resonance imaging for the non-invasive determination of cardiac output.J Clin Monit Comput. 2016 Aug;30(4):399-408. doi: 10.1007/s10877-015-9731-6. Epub 2015 Jun 27. J Clin Monit Comput. 2016. PMID: 26115774
-
Cardiac index measurements by transcutaneous Doppler ultrasound and transthoracic echocardiography in adult and pediatric emergency patients.J Clin Monit Comput. 2010 Jun;24(3):237-47. doi: 10.1007/s10877-010-9240-6. Epub 2010 Jun 20. J Clin Monit Comput. 2010. PMID: 20563629
-
Authors' Response to a Letter to the Editor.Pediatr Cardiol. 2021 Apr;42(4):983-984. doi: 10.1007/s00246-021-02605-1. Epub 2021 Apr 9. Pediatr Cardiol. 2021. PMID: 33835231 No abstract available.
-
Ultrasound Cardiac Output Monitor (USCOM™) Measurements Prove Unreliable Compared to Cardiac Magnetic Resonance Imaging in Adolescents with Cardiac Disease.Pediatr Cardiol. 2021 Mar;42(3):692-699. doi: 10.1007/s00246-020-02531-8. Epub 2021 Jan 4. Pediatr Cardiol. 2021. PMID: 33394105
-
Third-Trimester Cardiovascular Function and Risk of Hypertensive Disorders of Pregnancy.J Am Heart Assoc. 2024 Oct 15;13(20):e032673. doi: 10.1161/JAHA.123.032673. Epub 2024 Oct 11. J Am Heart Assoc. 2024. PMID: 39392169 Free PMC article.
References
-
- Connors AF Jr, Speroff T, Dawson NV, et al. The effectiveness of right heart catheterization in the initial care of critically ill patients. SUPPORT Investigators. JAMA 1996; 276: 889–897. - PubMed
-
- Pohost GM, Hung L, Doyle M. Clinical use of cardiovascular magnetic resonance. Circulation 2003; 108: 647–653. - PubMed
-
- Appel PL, Kram HB, Mackabee J, Fleming AW, Shoemaker WC. Comparison of measurements of cardiac output by bioimpedance and thermodilution in severely ill surgical patients. Crit Care Med 1986; 14: 933–935. - PubMed
-
- Franko E, Van De Water J, Wang X. Ideal measurement of cardiac output: is impedance cardiography the answer? Vasc Endovascular Surg 1991; 25: 550–558.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical