Mass spectrometer for respiratory research
- PMID: 20045084
- DOI: 10.1016/j.resp.2009.12.013
Mass spectrometer for respiratory research
Abstract
The mass spectrometer is the ideal gas analyser for many applications in respiratory research. A number of companies which manufactured mass spectrometers suitable for use in respiratory research have long since ceased production. Two modifications in the presently manufactured MAX300-LG (Extrel, Pittsburgh, PA, USA) have made it suitable for studying respiratory physiology. The initial step was to bypass the four-port valve and connect the sampling capillary directly to the roughing pump. The second step was to replace the original capillary, which had an internal diameter of 0.4mm, by a 2-m long capillary having an internal diameter of 0.15 mm. The 90% response time was shortened to 85 ms, which is suitable for respiratory research.
Copyright 2009 Elsevier B.V. All rights reserved.