Safety and function of a new clinical intracerebral microinjection instrument for stem cells and therapeutics examined in the Göttingen minipig
- PMID: 20051711
- DOI: 10.1159/000268743
Safety and function of a new clinical intracerebral microinjection instrument for stem cells and therapeutics examined in the Göttingen minipig
Abstract
Background: A new intracerebral microinjection instrument (IMI) allowing multiple electrophysiologically guided microvolume injections from a single proximal injection path in rats has been adapted to clinical use by coupling the IMI to an FHC microTargeting Manual Drive, designed to be used with standard stereotactic frame-based systems and FHC frameless microTargeting Platforms.
Methods: The function and safety of the device was tested by conducting bilateral electrophysiologically guided microinjections of fluorescent microspheres in the substantia nigra of 4 Göttingen minipigs.
Results: The device was easy to handle and enabled accurate electrophysiologically guided targeting of the substantia nigra with minimal local tissue damage.
Conclusion: The IMI is suitable for clinical use and may prove useful for various stereotactic procedures that require high levels of precision and/or three-dimensional distribution of therapeutics within the brain.
2009 S. Karger AG, Basel.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical