Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
Comparative Study
. 2010 Jun;51(6):3184-92.
doi: 10.1167/iovs.09-4886. Epub 2010 Jan 6.

Monoscopic versus stereoscopic retinal photography for grading diabetic retinopathy severity

Affiliations
Comparative Study

Monoscopic versus stereoscopic retinal photography for grading diabetic retinopathy severity

Helen K Li et al. Invest Ophthalmol Vis Sci. 2010 Jun.

Abstract

Purpose: To assess agreement between monoscopic and stereoscopic photography for research classification of the severity of diabetic retinopathy (DR).

Methods: Monoscopic digital (MD) images were compared with stereo digital (SD) and film (SF) photographs from a 152-eye cohort with full-spectrum Early Treatment Diabetic Retinopathy Study (ETDRS) severity levels for agreement on severity level, DR presence with ascending severity threshold, presence of DR index lesions, and repeatability of grading.

Results: There was substantial agreement classifying ETDRS DR severity levels between MD and SF (kappa = 0.65, kappa(w) [linear weighted] = 0.87), MD and SD (kappa = 0.66, kappa(w) = 0.87), and SD and SF (kappa = 0.62, kappa(w) = 0.86) images. Marginal homogeneity analyses found no significant difference between MD and SF images (P = 0.53, Bhapkar test). The kappa agreement between MD and SF ranged from 0.80 to 0.94 for the presence or absence of eight ascending DR severity thresholds. Repeatability between the readers of the MD images was equal to or better than that of the readers of SD or SF images. Severity threshold grading repeatability between readers was similar with the MD and SF images. The kappa agreement between MD and SF for identifying diabetic retinopathy lesions ranged from moderate to almost perfect. The kappa comparisons showed that performance of grading new vessels on the disc in MD images was slightly lower than that with the SF images.

Conclusions: Monoscopic photography can equal the reliability of stereo photography for full ETDRS DR severity scale grading.