Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
Randomized Controlled Trial
. 2010 Aug;54(7):848-54.
doi: 10.1111/j.1399-6576.2009.02194.x. Epub 2010 Jan 6.

Exposure to anaesthetic trace gases during general anaesthesia: CobraPLA vs. LMA classic

Affiliations
Randomized Controlled Trial

Exposure to anaesthetic trace gases during general anaesthesia: CobraPLA vs. LMA classic

K Schebesta et al. Acta Anaesthesiol Scand. 2010 Aug.

Abstract

Background: To prospectively investigate the performance, sealing capacity and operating room (OR) staff exposure to waste anaesthetic gases during the use of the Cobra perilaryngeal airway (CobraPLA) compared with the laryngeal mask airway classic (LMA).

Methods: Sixty patients were randomly assigned to the CobraPLA or the LMA group. Insertion time, number of insertion attempts and airway leak pressures were assessed after induction of anaesthesia. Occupational exposure to nitrous oxide (N(2)O) and Sevoflurane (SEV) was measured at the anaesthetists' breathing zone and the patients' mouth using a photoacoustic infrared spectrometer.

Results: N(2)O waste gas concentrations differed significantly in the anaesthetist's breathing zone (11.7+/-7.2 p.p.m. in CobraPLA vs. 4.1+/-4.3 p.p.m. in LMA, P=0.03), whereas no difference could be shown in SEV concentrations. Correct CobraPLA positioning was possible in 28 out of 30 patients (more than one attempt necessary in five patients). Correct positioning of the LMA classic was possible in all 30 patients (more than one attempt in three patients). Peak airway pressure was higher in the CobraPLA group (16+/-3 vs. 14+/-2 cmH(2)O, P=0.01). The average leak pressure of the CobraPLA was 24+/-4 cmH(2)O, compared with 20+/-4 cmH(2)O of the LMA classic (P<0.001; all values means+/-SD).

Conclusion: Despite higher airway seal pressures, the CobraPLA caused higher intraoperative N(2)O trace concentrations in the anaesthetists' breathing zone.

PubMed Disclaimer

Similar articles

Cited by

Publication types

LinkOut - more resources