Device-based impedance measurement is a useful and accurate tool for direct assessment of intrathoracic fluid accumulation in heart failure
- PMID: 20056595
- DOI: 10.1093/europace/eup413
Device-based impedance measurement is a useful and accurate tool for direct assessment of intrathoracic fluid accumulation in heart failure
Abstract
Aims: Heart failure patients are often equipped with implanted devices and are frequently hospitalized due to volume overload. Reliable prediction of imminent fluid congestion has the potential to provide early detection of cardiac decompensation and therefore might be capable of enhancing therapy management. We investigated whether implant-based impedance (Z) measurement is closely correlated with directly assessed extravascular lung water and might thus be useful for patient monitoring.
Methods and results: In sheep, pulmonary fluid congestion was induced. Continuous haemodynamic monitoring was performed and extravascular lung water index (EVLWI) assessed. An implanted device with a right ventricular lead measured Z using different electrode configurations. All animals developed gradual pulmonary fluid accumulation leading to inclining lung oedema: EVLWI did increase from 9.5 +/- 1 to 21.1 +/- 5.1 mL/kg (+127%). A concomitant decrease of Z by up to 23%, depending on the electrode configuration, was observed and regression analysis between Z and EVLWI yielded a significant inverse correlation.
Conclusion: Changes of Z show a strong inverse correlation with changes of directly measured EVLWI. This allows the application of Z as a measure of intrathoracic fluid status and has the potential to optimize patient care, especially in the context of evolving telemedicine concepts.
Comment in
-
Overcoming the resistance to impedance monitoring in heart failure patients.Europace. 2010 May;12(5):614-5. doi: 10.1093/europace/euq055. Epub 2010 Mar 10. Europace. 2010. PMID: 20219751 No abstract available.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Medical