Pitfalls of liver stiffness measurement: a 5-year prospective study of 13,369 examinations
- PMID: 20063276
- DOI: 10.1002/hep.23425
Pitfalls of liver stiffness measurement: a 5-year prospective study of 13,369 examinations
Abstract
Liver stiffness measurement (LSM) based on transient elastography (TE, FibroScan) is gaining in popularity for noninvasive assessment of liver fibrosis. However, LSM has limitations, which have not yet been thoroughly evaluated. We prospectively investigated the frequency and determinants of LSM failure and unreliable results over a 5-year period, based on 13,369 examinations (134,239 shots). LSM failure was defined as zero valid shots, and unreliable examinations were defined as fewer than 10 valid shots, an interquartile range (IQR)/LSM greater than 30%, or a success rate less than 60%. LSM failure occurred in 3.1% of all examinations (4% at first examination [n = 7261]) and was independently associated at first examination with body mass index (BMI) greater than 30 kg/m(2) (odds ratio [OR], 7.5; 95% confidence interval [CI], 5.6-10.2; P = 0.0001), operator experience fewer than 500 examinations (OR 2.5 [1.6-4.0]; P = 0.0001); age greater than 52 years (OR 2.3 [1.6-3.2]; P = 0.0001), and type 2 diabetes (OR 1.6 [1.1-2.2]; P = 0.009). Unreliable results were obtained in a further 15.8% of cases (17% at first examination) and were independently associated at first examination with BMI greater than 30 kg/m(2) (OR 3.3 [2.8-4.0]; P = 0.0001), operator experience fewer than 500 examinations (OR 3.1 [2.4-3.9]; P = 0.0001), age greater than 52 years (OR 1.8 [1.6-2.1]; P = 0.0001), female sex (OR 1.4 [1.2-1.6], P = 0.0001), hypertension (OR 1.3 [1.1-1.5]; P = 0.003), and type 2 diabetes (OR 1.2 [1.0-1.5]; P = 0.05). When metabolic syndrome and waist circumference were taken into account in a subgroup of 2835 patients, waist circumference was the most important determinant of LSM failure and unreliable results.
Conclusion: In our experience, liver stiffness measurements are uninterpretable in nearly one in five cases. The principal reasons are obesity, particularly increased waist circumference, and limited operator experience. These results emphasize the need for adequate operator training and for technological improvements in specific patient subpopulations.
Similar articles
-
Feasibility and diagnostic performance of the FibroScan XL probe for liver stiffness measurement in overweight and obese patients.Hepatology. 2012 Jan;55(1):199-208. doi: 10.1002/hep.24624. Epub 2011 Nov 18. Hepatology. 2012. PMID: 21898479 Clinical Trial.
-
Longer duration of transient elastography predicts unreliable liver stiffness measurements.Eur J Gastroenterol Hepatol. 2015 Jun;27(6):655-9. doi: 10.1097/MEG.0000000000000310. Eur J Gastroenterol Hepatol. 2015. PMID: 25882128
-
The feasibility and reliability of transient elastography using Fibroscan®: a practice audit of 2335 examinations.Can J Gastroenterol Hepatol. 2014 Mar;28(3):143-9. doi: 10.1155/2014/952684. Can J Gastroenterol Hepatol. 2014. PMID: 24619636 Free PMC article.
-
The frequency and determinants of liver stiffness measurement failure: a retrospective study of "real-life" 38,464 examinations.PLoS One. 2014 Aug 14;9(8):e105183. doi: 10.1371/journal.pone.0105183. eCollection 2014. PLoS One. 2014. PMID: 25122123 Free PMC article.
-
Range of Normal Liver Stiffness and Factors Associated With Increased Stiffness Measurements in Apparently Healthy Individuals.Clin Gastroenterol Hepatol. 2019 Jan;17(1):54-64.e1. doi: 10.1016/j.cgh.2018.08.069. Epub 2018 Sep 7. Clin Gastroenterol Hepatol. 2019. PMID: 30196155
Cited by
-
Identifying High-Risk NASH Patients: What We Know so Far.Hepat Med. 2020 Aug 21;12:125-138. doi: 10.2147/HMER.S265473. eCollection 2020. Hepat Med. 2020. PMID: 32982495 Free PMC article. Review.
-
Task-based optimization of flip angle for fibrosis detection in T1-weighted MRI of liver.J Med Imaging (Bellingham). 2016 Jul;3(3):035502. doi: 10.1117/1.JMI.3.3.035502. Epub 2016 Jul 21. J Med Imaging (Bellingham). 2016. PMID: 27446971 Free PMC article.
-
NIKEI: a new inexpensive and non-invasive scoring system to exclude advanced fibrosis in patients with NAFLD.PLoS One. 2013;8(3):e58360. doi: 10.1371/journal.pone.0058360. Epub 2013 Mar 26. PLoS One. 2013. PMID: 23555578 Free PMC article. Clinical Trial.
-
Real-time elastography as a noninvasive assessment of liver fibrosis in chronic hepatitis C Egyptian patients: a prospective study.Ann Gastroenterol. 2016 Jul-Sep;29(3):358-62. doi: 10.20524/aog.2016.0033. Epub 2016 Apr 19. Ann Gastroenterol. 2016. PMID: 27366038 Free PMC article.
-
JSUM ultrasound elastography practice guidelines: liver.J Med Ultrason (2001). 2013 Oct;40(4):325-57. doi: 10.1007/s10396-013-0460-5. Epub 2013 Aug 16. J Med Ultrason (2001). 2013. PMID: 27277450 Review.
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical