Influence of surgical technique and surface roughness on the primary stability of an implant in artificial bone with different cortical thickness: a laboratory study
- PMID: 20070754
- DOI: 10.1111/j.1600-0501.2009.01823.x
Influence of surgical technique and surface roughness on the primary stability of an implant in artificial bone with different cortical thickness: a laboratory study
Abstract
Objective: The aim of this biomechanical study was to assess the interrelated effect of both surface roughness and surgical technique on the primary stability of dental implants.
Material and methods: For the experiment, 160 screw-designed implants (Biocomp), with either a machined or an etched surface topography, were inserted into polyurethane foam blocks (Sawbones). As an equivalent of trabecular bone, a density of 0.48 g/cm(3) was chosen. To mimic the cortical layer, on top of these blocks short-fibre-filled epoxy sheets were attached with a thickness varying from 0 to 2.5 mm. The implant sites were prepared using either a press-fit or an undersized technique. To measure the primary stability of the implant, both the insertion and the removal torques were scored.
Results: Independent of the surgical technique used, both implant types showed an increased insertion and removal torque values with increasing cortical thickness, although >2 mm cortical layer no further increase in insertion torque was observed. In the models with only trabecular bone (without cortical layer) and with a 1 mm cortical layer, both implant types showed a statistically higher insertion and removal torque values for undersized compared with the press-fit technique. In addition, etched implants showed a statistically higher insertion and removal torque mean values compared with machined implants. In the models with 2 and 2.5 mm cortical layers, with respect to the insertion torque values, no effect of either implantation technique or implant surface topography could be observed.
Conclusion: The placement of etched implants in synthetic bone models using an undersized preparation technique resulted in enhanced primary implant stability. A correlation was found between the primary stability and the cortical thickness. However, at or above a cortical thickness of 2 mm, the effect of both an undersized surgical approach, as also the presence of a roughened (etched) implant surface, had no extra effect. Besides the mechanical aspects, the biological effect of undersized drilling, i.e. the bone response on the extra insertion torque forces should also be elucidated. Therefore, additional in vivo studies are needed.
Similar articles
-
Influence of the surgical technique and surface roughness on the primary stability of an implant in artificial bone with a density equivalent to maxillary bone: a laboratory study.Clin Oral Implants Res. 2009 Apr;20(4):327-32. doi: 10.1111/j.1600-0501.2008.01692.x. Clin Oral Implants Res. 2009. PMID: 19298286
-
The effects of implant surface roughness and surgical technique on implant fixation in an in vitro model.Clin Oral Implants Res. 2006 Apr;17(2):172-8. doi: 10.1111/j.1600-0501.2005.01202.x. Clin Oral Implants Res. 2006. PMID: 16584413
-
Primary stability of turned and acid-etched screw-type implants: a removal torque and histomorphometric study in rabbits.Int J Oral Maxillofac Implants. 2007 Nov-Dec;22(6):886-92. Int J Oral Maxillofac Implants. 2007. PMID: 18271369
-
The role of primary stability for successful immediate loading of dental implants. A literature review.J Dent. 2010 Aug;38(8):612-20. doi: 10.1016/j.jdent.2010.05.013. Epub 2010 Jun 11. J Dent. 2010. PMID: 20546821 Review.
-
Implant Insertion Torque: Its Role in Achieving Primary Stability of Restorable Dental Implants.Compend Contin Educ Dent. 2017 Feb;38(2):88-95; quiz 96. Compend Contin Educ Dent. 2017. PMID: 28156122 Review.
Cited by
-
The Effect of Coronal Implant Design and Drilling Protocol on Bone-to-Implant Contact: A 3-Month Study in the Minipig Calvarium.Materials (Basel). 2021 May 18;14(10):2645. doi: 10.3390/ma14102645. Materials (Basel). 2021. PMID: 34070127 Free PMC article.
-
Primary Stability Optimization by Using Fixtures with Different Thread Depth According To Bone Density: A Clinical Prospective Study on Early Loaded Implants.Materials (Basel). 2019 Jul 27;12(15):2398. doi: 10.3390/ma12152398. Materials (Basel). 2019. PMID: 31357620 Free PMC article.
-
Influence of different surgical techniques on primary implant stability in the posterior maxilla: a randomized controlled clinical trial.Clin Oral Investig. 2023 Jul;27(7):3499-3508. doi: 10.1007/s00784-023-04962-y. Epub 2023 Mar 28. Clin Oral Investig. 2023. PMID: 36977760 Free PMC article. Clinical Trial.
-
Evaluation of Bone Density for Primary Implant Stability Using a Newly Designed Drill: An In Vitro Study on Polyurethane Bone Blocks.Clin Exp Dent Res. 2024 Dec;10(6):e70048. doi: 10.1002/cre2.70048. Clin Exp Dent Res. 2024. PMID: 39610027 Free PMC article.
-
Wettability and adhesion of nanotubes applied to the surface of titanium implants by anodic oxidation.Braz Oral Res. 2024 Sep 9;38:e091. doi: 10.1590/1807-3107bor-2024.vol38.0091. eCollection 2024. Braz Oral Res. 2024. PMID: 39258602 Free PMC article.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources
Research Materials