Double-spin-echo diffusion weighting with a modified eddy current adjustment
- PMID: 20071120
- DOI: 10.1016/j.mri.2009.12.004
Double-spin-echo diffusion weighting with a modified eddy current adjustment
Abstract
Magnetic field inhomogeneities like eddy current-related gradient fields cause geometric distortions in echo-planar imaging (EPI). This in particular affects diffusion-weighted imaging where these distortions vary with the direction of the diffusion weighting and hamper the accurate determination of diffusion parameters. The double-spin-echo preparation often used aims to reduce the cumulative eddy current effect by adjusting the diffusion-weighting gradient pulse durations to the time constant of the dominant eddy current contribution. However, eddy currents with a variety of time constants may be present and cause residual distortions. Here, a modification is proposed where the two bipolar gradient pairs of the preparation are adjusted independently to different time constants. At the expense of a slightly prolonged echo time, residual geometric distortions and correspondingly increased values of the diffusion anisotropy can be reduced as is demonstrated in phantoms and the human brain. Thus, it may help to improve the reliability of diffusion-weighted EPI.
Copyright 2010 Elsevier Inc. All rights reserved.
Similar articles
-
Eddy-current compensated diffusion weighting with a single refocusing RF pulse.Magn Reson Med. 2009 Mar;61(3):748-54. doi: 10.1002/mrm.21899. Magn Reson Med. 2009. PMID: 19132755
-
Generalized MAGSTE with bipolar diffusion-weighting gradient pulses.J Magn Reson. 2009 Aug;199(2):214-24. doi: 10.1016/j.jmr.2009.05.004. Epub 2009 May 13. J Magn Reson. 2009. PMID: 19524465
-
Diffusion-weighted imaging in patients with acute brain ischemia at 3 T: current possibilities and future perspectives comparing conventional echoplanar diffusion-weighted imaging and fast spin echo diffusion-weighted imaging sequences using BLADE (PROPELLER).Invest Radiol. 2009 Jun;44(6):351-9. doi: 10.1097/RLI.0b013e3181a00d09. Invest Radiol. 2009. PMID: 19363447
-
Spin echo magnetic resonance imaging.J Magn Reson Imaging. 2013 Apr;37(4):805-17. doi: 10.1002/jmri.24068. J Magn Reson Imaging. 2013. PMID: 23526758 Review.
-
[Diffusion-weighted MR imaging of the spine and cord].J Radiol. 2010 Mar;91(3 Pt 2):352-66; quiz 367-8. doi: 10.1016/s0221-0363(10)70051-2. J Radiol. 2010. PMID: 20508570 Review. French.
Cited by
-
Oscillating diffusion-encoding with a high gradient-amplitude and high slew-rate head-only gradient for human brain imaging.Magn Reson Med. 2020 Aug;84(2):950-965. doi: 10.1002/mrm.28180. Epub 2020 Feb 3. Magn Reson Med. 2020. PMID: 32011027 Free PMC article.
-
Denoising and Multiple Tissue Compartment Visualization of Multi-b-Valued Breast Diffusion MRI.J Magn Reson Imaging. 2021 Jan;53(1):271-282. doi: 10.1002/jmri.27268. Epub 2020 Jul 2. J Magn Reson Imaging. 2021. PMID: 32614125 Free PMC article.
-
Comparison of twice refocused spin echo versus stimulated echo diffusion tensor imaging for tracking muscle fibers.J Magn Reson Imaging. 2015 Mar;41(3):624-32. doi: 10.1002/jmri.24585. Epub 2014 Feb 19. J Magn Reson Imaging. 2015. PMID: 24554376 Free PMC article.
-
Dynamic correction of artifacts due to susceptibility effects and time-varying eddy currents in diffusion tensor imaging.Neuroimage. 2011 Aug 15;57(4):1343-7. doi: 10.1016/j.neuroimage.2011.06.008. Epub 2011 Jun 14. Neuroimage. 2011. PMID: 21689763 Free PMC article.
-
Pre-excitation gradients for eddy current nulled convex optimized diffusion encoding (Pre-ENCODE).Magn Reson Med. 2024 Aug;92(2):573-585. doi: 10.1002/mrm.30068. Epub 2024 Mar 19. Magn Reson Med. 2024. PMID: 38501914 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources