Osseointegration of one-piece zirconia implants compared with a titanium implant of identical design: a histomorphometric study in the dog
- PMID: 20074240
- DOI: 10.1111/j.1600-0501.2009.01832.x
Osseointegration of one-piece zirconia implants compared with a titanium implant of identical design: a histomorphometric study in the dog
Abstract
Objective: The aim of this study was to evaluate osseointegration of one-piece zirconia vs. titanium implants depending on their insertion depth by histomorphometry.
Material and methods: Four one-piece implants of identical geometry were inserted on each side of six mongrel dogs: (1) an uncoated zirconia implant, (2) a zirconia implant coated with a calcium-liberating titanium oxide coating, (3) a titanium implant and (4) an experimental implant made of a synthetic material (polyetheretherketone). In a split-mouth manner they were inserted in submerged and non-submerged gingival healing modes. After 4 months, dissected blocks were stained with toluidine blue in order to histologically assess the bone-to-implant contact (BIC) rates and the bone levels (BL) of the implants.
Results: All 48 implants were osseointegrated clinically and histologically. Histomorphometrically, BL in the crestal implant part did not differ significantly with regard to material type or healing modality. The submerged coated zirconia implants tended to offer the most stable crestal BL. The histometric results reflected the different healing modes by establishing different BL. The median BIC of the apical implant part of the zirconia and titanium group amounted to 59.2% for uncoated zirconia, 58.3% for coated zirconia, 26.8% for the synthetic material and 41.2% for titanium implants.
Conclusions: Within the limits of this animal study, it is concluded that zirconia implants are capable of establishing close BIC rates similar to what is known from the osseointegration behaviour of titanium implants with the same surface modification and roughness.
Similar articles
-
Soft tissue healing at one-piece zirconia implants compared to titanium and PEEK implants of identical design: a histomorphometric study in the dog.Int J Periodontics Restorative Dent. 2013 Sep-Oct;33(5):669-77. doi: 10.11607/prd.1043. Int J Periodontics Restorative Dent. 2013. PMID: 23998163
-
Comparison of zirconia and titanium implants after a short healing period. A pilot study in minipigs.Int J Oral Maxillofac Surg. 2010 Jun;39(6):585-92. doi: 10.1016/j.ijom.2010.01.015. Epub 2010 Feb 20. Int J Oral Maxillofac Surg. 2010. PMID: 20172693
-
Biomechanical and histomorphometric comparison between zirconia implants with varying surface textures and a titanium implant in the maxilla of miniature pigs.Clin Oral Implants Res. 2007 Oct;18(5):662-8. doi: 10.1111/j.1600-0501.2007.01401.x. Epub 2007 Jun 30. Clin Oral Implants Res. 2007. PMID: 17608736
-
Zirconia dental implants: a literature review.J Oral Implantol. 2011 Jun;37(3):367-76. doi: 10.1563/AAID-JOI-D-09-00079. Epub 2010 Jun 14. J Oral Implantol. 2011. PMID: 20545529 Review.
-
Osseointegration of titanium, titanium alloy and zirconia dental implants: current knowledge and open questions.Periodontol 2000. 2017 Feb;73(1):22-40. doi: 10.1111/prd.12179. Periodontol 2000. 2017. PMID: 28000277 Review.
Cited by
-
Finite element evaluation to compare stress pattern in bone surrounding implant with carbon fiber-reinforced poly-ether-ether-ketone and commercially pure titanium implants.Natl J Maxillofac Surg. 2022 May-Aug;13(2):243-247. doi: 10.4103/njms.njms_354_21. Epub 2022 Jul 15. Natl J Maxillofac Surg. 2022. PMID: 36051812 Free PMC article.
-
Biomechanical and histological evaluation of the osseointegration capacity of two types of zirconia implant.Int J Nanomedicine. 2016 Dec 7;11:6507-6516. doi: 10.2147/IJN.S119519. eCollection 2016. Int J Nanomedicine. 2016. PMID: 27994456 Free PMC article.
-
Biomaterials and Clinical Application of Dental Implants in Relation to Bone Density-A Narrative Review.J Clin Med. 2023 Nov 3;12(21):6924. doi: 10.3390/jcm12216924. J Clin Med. 2023. PMID: 37959389 Free PMC article. Review.
-
Fatigue and Fracture Resistance Testing of Polyether Ether Ketone (PEEK) Implant Abutments in an Ex Vivo Chewing Simulator Model.Materials (Basel). 2022 Oct 6;15(19):6927. doi: 10.3390/ma15196927. Materials (Basel). 2022. PMID: 36234281 Free PMC article.
-
Finite Element Analysis of the Stress and Deformation in Bone Caused by CFR-PEEK, Titanium, and Zirconia Ceramic Implants.J Pharm Bioallied Sci. 2024 Jul;16(Suppl 3):S2078-S2080. doi: 10.4103/jpbs.jpbs_45_24. Epub 2024 Jun 7. J Pharm Bioallied Sci. 2024. PMID: 39346482 Free PMC article.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources