Effect of joint laxity on polyethylene wear in total knee replacement
- PMID: 20074735
- DOI: 10.1016/j.jbiomech.2009.12.016
Effect of joint laxity on polyethylene wear in total knee replacement
Abstract
Experimental simulator studies are frequently performed to evaluate wear behavior in total knee replacement. It is vital that the simulation conditions match the physiological situation as closely as possible. To date, few experimental wear studies have examined the effects of joint laxity on wear and joint kinematics and the absence of the anterior cruciate ligament has not been sufficiently taken into account in simulator wear studies. The aim of this study was to investigate different ligament and soft tissue models with respect to wear and kinematics. A virtual soft tissue control system was used to simulate different motion restraints in a force-controlled knee wear simulator. The application of more realistic and sophisticated ligament models that considered the absence of anterior cruciate ligament lead to a significant increase in polyethylene wear (p=0.02) and joint kinematics (p<0.01). We recommend the use of more complex ligament models to appropriately simulate the function of the human knee joint and to evaluate the wear behavior of total knee replacements. A feasible simulation model is presented.
Copyright 2009 Elsevier Ltd. All rights reserved.
Similar articles
-
Evaluation of the effect of joint constraints on the in situ force distribution in the anterior cruciate ligament.J Orthop Res. 1997 Mar;15(2):278-84. doi: 10.1002/jor.1100150218. J Orthop Res. 1997. PMID: 9167632
-
Fixed and mobile bearing total knee arthroplasty--influence on wear generation, corresponding wear areas, knee kinematics and particle composition.Clin Biomech (Bristol). 2009 Feb;24(2):210-7. doi: 10.1016/j.clinbiomech.2008.11.006. Epub 2008 Dec 31. Clin Biomech (Bristol). 2009. PMID: 19118930
-
In vitro investigation of knee joint kinematics following cruciate retaining versus cruciate sacrificing total knee arthroplasty.Acta Orthop Belg. 2003 Oct;69(5):433-40. Acta Orthop Belg. 2003. PMID: 14648953
-
The dominance of cyclic sliding in producing wear in total knee replacements.Clin Orthop Relat Res. 1991 Dec;(273):253-60. Clin Orthop Relat Res. 1991. PMID: 1959278 Review.
-
Characterizing the motion of total knee replacements in laboratory tests.Clin Orthop Relat Res. 2003 May;(410):54-68. doi: 10.1097/01.blo.0000063599.67412.44. Clin Orthop Relat Res. 2003. PMID: 12771817 Review.
Cited by
-
[Principles of tribological analysis of endoprostheses].Orthopade. 2012 Oct;41(10):844-52. doi: 10.1007/s00132-012-1948-1. Orthopade. 2012. PMID: 23052851 German.
-
Wear in total knee arthroplasty--just a question of polyethylene?: Metal ion release in total knee arthroplasty.Int Orthop. 2014 Feb;38(2):335-40. doi: 10.1007/s00264-013-2162-4. Epub 2013 Nov 12. Int Orthop. 2014. PMID: 24218115 Free PMC article.
-
Wear behavior of an unstable knee: stabilization via implant design?Biomed Res Int. 2014;2014:821475. doi: 10.1155/2014/821475. Epub 2014 Sep 9. Biomed Res Int. 2014. PMID: 25276820 Free PMC article.
-
An Interesting Case of Osteolysis With Accompanying Metallosis in a Primary Total Knee Arthroplasty.Arthroplast Today. 2021 Aug 26;11:81-87. doi: 10.1016/j.artd.2021.07.002. eCollection 2021 Oct. Arthroplast Today. 2021. PMID: 34485655 Free PMC article.
-
Wear testing of moderate activities of daily living using in vivo measured knee joint loading.PLoS One. 2015 Mar 26;10(3):e0123155. doi: 10.1371/journal.pone.0123155. eCollection 2015. PLoS One. 2015. PMID: 25811996 Free PMC article.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Medical