Biomechanical evaluation of short-segment posterior instrumentation with and without crosslinks in a human cadaveric unstable thoracolumbar burst fracture model
- PMID: 20075769
- DOI: 10.1097/BRS.0b013e3181bda4e6
Biomechanical evaluation of short-segment posterior instrumentation with and without crosslinks in a human cadaveric unstable thoracolumbar burst fracture model
Abstract
Study design: This study evaluates the biomechanical characteristics of spinal instrumentation constructs in a human unstable thoracolumbar burst fracture model simulated by corpectomy.
Objective: To compare the biomechanical characteristics of short-segment posterior instrumentation, with and without crosslinks, in a human unstable burst fracture model simulated by corpectomy.
Summary of background data: Unstable thoracolumbar burst fractures are serious injuries, and their management remains controversial. Some authors advocate the use of short-segment posterior instrumentation for certain burst fractures. Whether crosslinks contribute additional stability has not been determined.
Methods: Six fresh frozen human spines (T10-L2) were potted to isolate the T11-L1 segments, and biomechanically tested in axial rotation, lateral bending, flexion, and extension. A custom spine testing system was used that allows motion with 6 degrees of freedom. After testing was completed on intact specimens, a corpectomy was performed at T12 to simulate an unstable burst fracture with loss of anterior and middle column support. Short-segment transpedicular instrumentation was then performed from T11 to L1. Each specimen was retested with 1, 2, or no crosslinks. Construct stiffness and motion data were analyzed with each intact specimen serving as its own internal control.
Results: Torsional stiffness in axial rotation was significantly increased (P < 0.05) in short-segment fixation constructs with 1 and 2 crosslinks, but none was restored to the preinjury baseline level. Significant reductions in standardized motion were also achieved with 1 and 2 crosslinks compared to no crosslinks (P < 0.05), but they remained greater than baseline. Crosslinks significantly increased stiffness and decreased motion in lateral bending, beyond the baseline level (P < 0.05). In flexion, all constructs had significantly decreased stiffness and increased motion compared to the intact specimen (P < 0.05), with crosslinks providing no additional benefit. Conversely, none of the constructs demonstrated a significant change in extension compared to baseline (P > 0.05). When attempting to load the constructs to failure, screw pullout was seen in all specimens.
Conclusion: Crosslinks, when added to short-segment posterior fixation, improve stiffness and decrease motion in axial rotation, but do not restore baseline stability in this corpectomy model. Short-segment posterior fixation is also inadequate in restoring stability in flexion with injuries of this severity. Short-segment posterior instrumentation alone can achieve baseline stability in lateral bending, and crosslinks provide even greater stiffness.
Similar articles
-
In vitro analysis of anterior and posterior fixation in an experimental unstable burst fracture model.J Spinal Disord Tech. 2008 May;21(3):216-24. doi: 10.1097/BSD.0b013e31807a2f61. J Spinal Disord Tech. 2008. PMID: 18458594
-
Biomechanical contribution of transverse connectors to segmental stability following long segment instrumentation with thoracic pedicle screws.Spine (Phila Pa 1976). 2008 Jul 1;33(15):E482-7. doi: 10.1097/BRS.0b013e31817c64d5. Spine (Phila Pa 1976). 2008. PMID: 18594445
-
Biomechanical effect of the extent of vertebral body fracture on the thoracolumbar spine with pedicle screw fixation: an in vitro study.J Clin Neurosci. 2008 Mar;15(3):286-90. doi: 10.1016/j.jocn.2006.12.007. Epub 2008 Jan 15. J Clin Neurosci. 2008. PMID: 18226530
-
The biomechanics of long versus short fixation for thoracolumbar spine fractures.Spine (Phila Pa 1976). 2006 May 15;31(11 Suppl):S70-9; discussion S104. doi: 10.1097/01.brs.0000218221.47230.dd. Spine (Phila Pa 1976). 2006. PMID: 16685240 Review.
-
[Effect of a crosslink or cerclage on the mechanical stability of an internal fixator].Orthopade. 1999 Aug;28(8):714-22. Orthopade. 1999. PMID: 10506374 Review. German.
Cited by
-
Does addition of crosslink to pedicle-screw-based instrumentation impact the development of the spinal canal in children younger than 5 years of age?Eur Spine J. 2015 Jul;24(7):1391-8. doi: 10.1007/s00586-014-3727-7. Epub 2014 Dec 20. Eur Spine J. 2015. PMID: 25527403
-
[Common pedicle screw placement under direct vision combined with dome shaped decompression via small incision for double segment thoracolumbar fracture with nerve injury].Zhongguo Xiu Fu Chong Jian Wai Ke Za Zhi. 2017 Jun 15;31(6):677-682. doi: 10.7507/1002-1892.201701092. Zhongguo Xiu Fu Chong Jian Wai Ke Za Zhi. 2017. PMID: 29798648 Free PMC article. Chinese.
-
Lumbar pedicle screw fixation with cortical bone trajectory: A review from anatomical and biomechanical standpoints.Spine Surg Relat Res. 2017 Nov 27;1(4):164-173. doi: 10.22603/ssrr.1.2017-0006. eCollection 2017. Spine Surg Relat Res. 2017. PMID: 31440629 Free PMC article. Review.
-
Cross-links in posterior pedicle screw-rod instrumentation of the spine: a systematic review on mechanical, biomechanical, numerical and clinical studies.Eur Spine J. 2021 Jan;30(1):34-49. doi: 10.1007/s00586-020-06597-z. Epub 2020 Oct 3. Eur Spine J. 2021. PMID: 33009949
-
Efficacy and Radiographic Analysis of Minimally Invasive Posterior Mono-Axial Pedicle Screw Fixation in Treating Thoracolumbar Burst Fractures.J Clin Med. 2022 Jan 20;11(3):516. doi: 10.3390/jcm11030516. J Clin Med. 2022. PMID: 35159967 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical
Research Materials