Comparison of the percutaneous screw placement precision of isocentric C-arm 3-dimensional fluoroscopy-navigated pedicle screw implantation and conventional fluoroscopy method with minimally invasive surgery
- PMID: 20075808
- DOI: 10.1097/BSD.0b013e31819877c8
Comparison of the percutaneous screw placement precision of isocentric C-arm 3-dimensional fluoroscopy-navigated pedicle screw implantation and conventional fluoroscopy method with minimally invasive surgery
Abstract
Study design: Retrospective clinical study to compare the percutaneous screw placement precision of isocentric C-arm (Iso-C) 3-dimensional fluoroscopy-navigated pedicle screw implantation and conventional fluoroscopy.
Objective: The purpose of this study was to evaluate the accuracy of clinical percutaneous pedicle screw placement (PPSP) using Iso-C 3-dimensional fluoroscopy navigation.
Summary of background data: The technique of PPSP recently has been applied to internal fixation for segmental lumbar instability as a form of minimally invasive spine surgery. The percutaneous insertion technique requires surgical skill and experience. The 3-dimensional fluoroscopy-based image-guidance navigation system has been proven to be effective in guiding accurate PPSP in the lumbar area of cadavers. However, there have been few clinical reports comparing the accuracy of Iso-C 3-dimensional fluoroscopy-navigated pedicle screw implantation and the conventional fluoroscopy method.
Methods: Here, 300 percutaneous pedicle screws were evaluated; half of them were inserted under Iso-C 3-dimensional navigation and the others under fluoroscopy. The accuracy of pedicle screw placement was examined postoperatively using an axial 2-mm slice computed tomography. Screw malpositioning was classified either as exposed screw (<50% of the screw outside the pedicle) or perforated pedicle (>50% of the screw outside the pedicle boundaries).
Results: Of the 150 pedicle screws placed with Iso-C 3-dimensional fluoroscopy-based image-guide assistance, 11 (7.3%) were classified as exposed screws and 0 (0%) as perforated pedicle. Of the 150 pedicle screws inserted under fluoroscopy, 18 (12%) were classified as exposed screw and 5 (3.3%) as perforated pedicle. The difference in frequency of screw misplacement between the procedure using Iso-C 3-dimensional fluoroscopy-based image-guide assistance and that using fluoroscopy was statistically significant (P<0.05; chi test).
Conclusions: This study demonstrates the feasibility of placing percutaneous posterior lumbar pedicle screws with the assistance of Iso-C 3-dimensional navigation.
Similar articles
-
Iso-C/3-dimensional neuronavigation versus conventional fluoroscopy for minimally invasive pedicle screw placement in lumbar fusion.Minim Invasive Neurosurg. 2010 Aug;53(4):184-90. doi: 10.1055/s-0030-1267926. Epub 2010 Dec 7. Minim Invasive Neurosurg. 2010. PMID: 21140333
-
Does Navigation Improve Pedicle Screw Placement Accuracy? Comparison Between Navigated and Non-navigated Percutaneous and Open Fixations.Acta Neurochir Suppl. 2017;124:289-295. doi: 10.1007/978-3-319-39546-3_42. Acta Neurochir Suppl. 2017. PMID: 28120086
-
Retrospective computed tomography scan analysis of percutaneously inserted pedicle screws for posterior transpedicular stabilization of the thoracic and lumbar spine: accuracy and complication rates.Spine (Phila Pa 1976). 2012 May 20;37(12):1092-100. doi: 10.1097/BRS.0b013e31823c80d8. Spine (Phila Pa 1976). 2012. PMID: 22037528
-
Accuracy of K-Wireless Insertion of Percutaneous Pedicle Screws Using Computer-Assisted Spinal Navigation: A Systematic Review and Single-Center Experience.World Neurosurg. 2020 Jun;138:e267-e274. doi: 10.1016/j.wneu.2020.02.079. Epub 2020 Feb 24. World Neurosurg. 2020. PMID: 32105880
-
The accuracy of pedicle screw placement using intraoperative image guidance systems.J Neurosurg Spine. 2014 Feb;20(2):196-203. doi: 10.3171/2013.11.SPINE13413. Epub 2013 Dec 20. J Neurosurg Spine. 2014. PMID: 24358998 Review.
Cited by
-
Percutaneous thoraco-lumbar-sacral pedicle screw placement accuracy results from a multi-center, prospective clinical study using a skin marker-based optical navigation system.Eur Spine J. 2022 Nov;31(11):3098-3108. doi: 10.1007/s00586-022-07387-5. Epub 2022 Sep 23. Eur Spine J. 2022. PMID: 36149493
-
Accuracy of Percutaneous Lumbosacral Pedicle Screw Placement Using the Oblique Fluoroscopic View Based on Computed Tomography Evaluations.Asian Spine J. 2016 Aug;10(4):630-8. doi: 10.4184/asj.2016.10.4.630. Epub 2016 Aug 16. Asian Spine J. 2016. PMID: 27559441 Free PMC article.
-
The evolution of image-guided lumbosacral spine surgery.Ann Transl Med. 2015 Apr;3(5):69. doi: 10.3978/j.issn.2305-5839.2015.02.01. Ann Transl Med. 2015. PMID: 25992368 Free PMC article. Review.
-
Low Radiation Protocol for Intraoperative Robotic C-Arm Can Enhance Adolescent Idiopathic Scoliosis Deformity Correction Accuracy and Safety.Global Spine J. 2024 Jun;14(5):1504-1514. doi: 10.1177/21925682221147867. Epub 2023 Jan 25. Global Spine J. 2024. PMID: 36695112 Free PMC article.
-
[Learning curve of minimally invasive pedicle screw placement].Surg Neurol Int. 2018 May 10;9(Suppl 2):S43-S49. doi: 10.4103/sni.sni_428_17. eCollection 2018. Surg Neurol Int. 2018. PMID: 29900033 Free PMC article. Spanish.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources