Comparative Study
. 2010 Sep;115(6):889-97.
doi: 10.1007/s11547-010-0504-5. Epub 2010 Jan 15.

Acoustic Radiation Force Impulse (ARFI) technique in ultrasound with Virtual Touch tissue quantification of the upper abdomen

[Article in English, Italian]
Affiliations
Comparative Study

Acoustic Radiation Force Impulse (ARFI) technique in ultrasound with Virtual Touch tissue quantification of the upper abdomen

[Article in English, Italian]
A Gallotti et al. Radiol Med. 2010 Sep.

Abstract

Purpose: Virtual Touch tissue quantification is an implementation of ultrasound (US) Acoustic Radiation Force Impulse (ARFI) imaging that provides numerical measurements (wave-velocity values) of tissue stiffness. The aim of this study was to define the normal values of shear-wave speed for the healthy liver, gallbladder, pancreas, spleen and kidneys.

Materials and methods: Thirty-five young healthy volunteers underwent Virtual Touch tissue quantification after having signed an informed consent form. All upper abdominal organs were examined by two independent operators. A phantom fluid model was also evaluated. All mean wave-velocity values were analysed and compared. Results. One hundred and forty measurements of liver, pancreas, spleen and kidneys, and 70 measurements of the gallbladder lumen were performed. Twenty measurements on the phantom were also performed. Comparing all measurements separately made by each operator in different parts of the organs, no statistically significant differences were observed. A "XXXX/0" value was always obtained from all measurements performed on the gallbladder lumen and on the phantom fluid model. Liver, pancreas, spleen and kidney mean values were 1.59 m/s, 1.40 m/s, 2.44 m/s and 2.24 m/s, respectively.

Conclusions: Virtual Touch tissue quantification is a new, promising implementation of the US ARFI technique, which provides numerical measurements of tissue stiffness. The mean shear-wave speed is lower in the pancreatic parenchyma than in the liver and kidney, whereas the spleen is characterised by the highest mean value. In simple fluids such as water, the value identified by the system with "XXXX" or 0, is always measured.

PubMed Disclaimer

Similar articles

Cited by

References

    1. Ultrason Imaging. 2007 Apr;29(2):87-104 - PubMed
    1. J Ultrasound Med. 2003 May;22(5):443-8 - PubMed
    1. Ultrason Imaging. 2008 Apr;30(2):95-111 - PubMed
    1. Ultrasound Med Biol. 2003 Dec;29(12):1715-23 - PubMed
    1. Rev Med Liege. 2007;62 Spec No:68-72 - PubMed

Publication types

LinkOut - more resources