. 2010 Jul;130(7):838-43.
doi: 10.3109/00016480903431139.

Validation study of portable device for the diagnosis of obstructive sleep apnea according to the new AASM scoring criteria: Watch-PAT 100

Affiliations

Validation study of portable device for the diagnosis of obstructive sleep apnea according to the new AASM scoring criteria: Watch-PAT 100

Ji Ho Choi et al. Acta Otolaryngol. 2010 Jul.

Abstract

Conclusion: Watch-PAT may provide an accurate and clinically effective portable monitoring method for the diagnosis of obstructive sleep apnea (OSA).

Objective: This study was designed to assess the accuracy and clinical efficacy of a wrist-worn portable device (Watch-PAT 100) to diagnose OSA.

Methods: Twenty-seven participants with suspected OSA underwent full polysomnography (PSG). Finally, 25 subjects successfully underwent portable monitoring using Watch-PAT after full PSG. The study population consisted of 21 males and 4 females, mean age 40.9 +/- 11.2 years (range 21-59). Mean body mass index (BMI) was 26.2 +/- 2.6 kg/m(2) (range 21.2-32.3). All PSG were manually scored according to the new scoring manual of the American Academy of Sleep Medicine (2007) and the Watch-PAT data were analyzed by the automatic algorithm.

Results: There was a high correlation of apnea-hypopnea index (AHI) (r = 0.94, p < 0.001) and lowest oxygen saturation (LSAT) (r = 0.90, p < 0.001) between the PSG and the Watch-PAT. A good agreement was also found between PSG AHI, PSG LSAT and PAT AHI, PAT LSAT, respectively. There was a significantly high concordance of the severity of AHI (Kendall tau-b = 0.897, p < 0.001) between the PSG and the Watch-PAT.

PubMed Disclaimer