Biomechanical comparison of 4 fixation techniques of sacral pedicle screw in osteoporotic condition

Affiliations
Comparative Study

Biomechanical comparison of 4 fixation techniques of sacral pedicle screw in osteoporotic condition

Bin-Sheng Yu et al. J Spinal Disord Tech. 2010 Aug.

Abstract

Study design: An in vitro biomechanical cadaver study.

Objectives: To compare the subsidence displacement after cyclic loading among 4 sacral pedicle screw fixations of bicortical, tricortical, standard polymethylmethacrylate (PMMA) augmentation, and sub-endplate PMMA augmentation in osteoporotic condition.

Summary of background data: Implant failure caused by screw loosening is a clinical problem for lumbosacral fusions, especially in osteoporotic patients. To improve sacral screw anchoring strength, the main fixation techniques need to be evaluated biomechanically.

Methods: For this study, 11 fresh osteoporotic cadaver sacra were harvested and bone mineral density was measured with dual-energy radiograph absorptiometry. A 7 mm diameter monoaxial pedicle screw (S1) was randomly assigned by side (left vs. right) and placed bicortically or tricortically. The 2 screws, followed 2000 cyclic compression loading of 30 to 250 N, were removed. The screw tracts were filled up with PMMA, then, screws 5 mm shorter than the bicortical or tricortical fixation were reinserted (defined as standard and sub-endplate PMMA augmented sacral screw fixations, respectively). The PMMA augmented screws were then retested as before. Screw subsidence displacement after 2000 cyclic loading was measured and compared.

Results: The average bone mineral density of 11 specimens was 0.71 g/cm, ranged from 0.65 to 0.78 g/cm. No significant difference of subsidence displacement was detected between tricortical and standard PMMA augmented screws (P>0.05), however, the 2 fixations exhibited markedly less subsidence than bicortical screw (P<0.05). Sub-endplate PMMA augmented screw showed the least subsidence among all the screws (P<0.05).

Conclusions: PMMA augmentation can increase the bonding strength of sacral screw-bone interface and the sub-endplate PMMA augmented sacral screw could obtain the highest stability among the 4 fixation techniques in osteoporotic condition.