Cardiac output derived from arterial pressure waveform analysis without calibration vs. thermodilution in septic shock: evolving accuracy of software versions
- PMID: 20090539
- DOI: 10.1097/EJA.0b013e3283333a92
Cardiac output derived from arterial pressure waveform analysis without calibration vs. thermodilution in septic shock: evolving accuracy of software versions
Abstract
Background and objective: We studied the evolution of software in the accuracy of the FloTrac/Vigileo system to measure cardiac output less invasively from arterial pressure waveform analysis without calibration, in comparison with pulmonary artery catheter-derived thermodilution measurements, in patients with septic shock and presumed alterations in vascular tone.
Methods: Nine patients who received a pulmonary artery catheter and were on mechanical ventilation and in sinus rhythm were monitored by the FloTrac/Vigileo. Paired cardiac output measurements by both techniques were analysed for 86 measurements in four patients using the 1.07 software version and 73 measurements in five subsequent patients using the later 1.10 version.
Results: For the 1.07 version, bias was -1.6 L min, precision 1.6 L min, limits of agreement -4.8-1.5 L min and error 48%. Measurements correlated at partial r equal to 0.32 (P = 0.003). For the 1.10 version, bias was -1.2 L min, precision 1.1 L min, limits of agreement -3.5-1.0 L min and error 32%. Measurements correlated at partial r equal to 0.90 (P < 0.001 vs. version 1.07). Differences were inversely related to mean cardiac output (P < 0.001, generalized estimating equations), particularly for software version 1.07 vs. 1.10 (P = 0.017, generalized estimating equation). Changes in thermodilution cardiac output over the course of time were also better tracked by the FloTrac/Vigileo when applying the latest software (P < 0.001, generalized estimating equation).
Conclusions: Evolving software versions are thus better able to account for the effect of vascular tone on cardiac output measurements by less invasive waveform analyses without calibration (FloTrac/Vigileo), so that the latter may become useful in the haemodynamic monitoring of septic shock.
Similar articles
-
Cardiac output derived from arterial pressure waveform analysis: validation of the third-generation software in patients undergoing orthotopic liver transplantation.Transplant Proc. 2012 Mar;44(2):433-7. doi: 10.1016/j.transproceed.2011.12.045. Transplant Proc. 2012. PMID: 22410036
-
Calibrated versus uncalibrated arterial pressure waveform analysis in monitoring cardiac output with transpulmonary thermodilution in patients with severe sepsis and septic shock: an observational study.Eur J Anaesthesiol. 2015 Jan;32(1):5-12. doi: 10.1097/EJA.0000000000000173. Eur J Anaesthesiol. 2015. PMID: 25387298
-
Cardiac output measured by a new arterial pressure waveform analysis method without calibration compared with thermodilution after cardiac surgery.J Cardiothorac Vasc Anesth. 2007 Oct;21(5):632-5. doi: 10.1053/j.jvca.2007.01.001. Epub 2007 Apr 5. J Cardiothorac Vasc Anesth. 2007. PMID: 17905265
-
Cardiac output derived from arterial pressure waveform.Curr Opin Anaesthesiol. 2009 Dec;22(6):804-8. doi: 10.1097/ACO.0b013e328332a473. Curr Opin Anaesthesiol. 2009. PMID: 19773648 Review.
-
[Arterial pressure-based cardiac output monitoring: 1. FloTrac sensor and SVV].Masui. 2009 Jul;58(7):838-47. Masui. 2009. PMID: 19618824 Review. Japanese.
Cited by
-
Cardiac output monitors in septic shock: do they deliver what matters? A systematic review and meta-analysis.Crit Care. 2025 Jul 12;29(1):299. doi: 10.1186/s13054-025-05547-9. Crit Care. 2025. PMID: 40652247 Free PMC article. Review.
-
The impact of systemic vascular resistance on the accuracy of the FloTrac/Vigileo™ system in the perioperative period of cardiac surgery: a prospective observational comparison study.J Clin Monit Comput. 2013 Dec;27(6):639-46. doi: 10.1007/s10877-013-9481-2. Epub 2013 Jun 8. J Clin Monit Comput. 2013. PMID: 23748600
-
Goal-directed fluid therapy using stroke volume variation on length of stay and postoperative gastrointestinal function after major abdominal surgery-a randomized controlled trial.BMC Anesthesiol. 2023 Dec 4;23(1):397. doi: 10.1186/s12871-023-02360-1. BMC Anesthesiol. 2023. PMID: 38049713 Free PMC article. Clinical Trial.
-
Cardiac output measured by uncalibrated arterial pressure waveform analysis by recently released software version 3.02 versus thermodilution in septic shock.J Clin Monit Comput. 2013 Apr;27(2):171-7. doi: 10.1007/s10877-012-9410-9. Epub 2012 Nov 15. J Clin Monit Comput. 2013. PMID: 23154918
-
Early non-invasive cardiac output monitoring in hemodynamically unstable intensive care patients: a multi-center randomized controlled trial.Crit Care. 2011 Jun 15;15(3):R148. doi: 10.1186/cc10273. Crit Care. 2011. PMID: 21676229 Free PMC article. Clinical Trial.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources