. 2010 Feb;194(2):398-406.
doi: 10.2214/AJR.09.2583.

Computer-aided volumetry of pulmonary nodules exhibiting ground-glass opacity at MDCT

Affiliations

Computer-aided volumetry of pulmonary nodules exhibiting ground-glass opacity at MDCT

Seitaro Oda et al. AJR Am J Roentgenol. 2010 Feb.

Abstract

Objective: The purpose of this study was to investigate the accuracy and reproducibility of results acquired with computer-aided volumetry software during MDCT of pulmonary nodules exhibiting ground-glass opacity.

Materials and methods: To evaluate the accuracy of computer-aided volumetry software, we performed thin-section helical CT of a chest phantom that included simulated 3-, 5-, 8-, 10-, and 12-mm-diameter ground-glass opacity nodules with attenuation of -800, -630, and -450 HU. Three radiologists measured the volume of the nodules and calculated the relative volume measurement error, which was defined as follows: (measured nodule volume minus assumed nodule volume / assumed nodule volume) x 100. Two radiologists performed two independent measurements of 59 nodules in humans. Intraobserver and interobserver agreement was evaluated with Bland-Altman methods.

Results: The relative volume measurement error for simulated ground-glass opacity nodules measuring 3 mm ranged from 51.1% to 85.2% and for nodules measuring 5 mm or more in diameter ranged from -4.1% to 7.1%. In the clinical study, for intraobserver agreement, the 95% limits of agreement were -14.9% and -13.7% and -16.6% to 15.7% for observers A and B. For interobserver agreement, these values were -16.3% to 23.7% for nodules 8 mm in diameter or larger.

Conclusion: With computer-aided volumetry of ground-glass opacity nodules, the relative volume measurement error was small for nodules 5 mm in diameter or larger. Intraobserver and interobserver agreement was relatively high for nodules 8 mm in diameter or larger.