Contamination of laryngoscope handles
- PMID: 20096953
- DOI: 10.1016/j.jhin.2009.09.015
Contamination of laryngoscope handles
Abstract
Despite use of sterile or disposable laryngoscope blades for each patient, disinfection of laryngoscope handles does not routinely occur, and these devices present a potential route of transmission of pathogens between patients and staff. A total of 192 specimens from 64 laryngoscope handles deemed 'ready for patient use' in the anaesthetic rooms of 32 operating theatres were semiquantitatively assessed for bacterial contamination. A further 116 specimens from 58 of the handles were tested for occult blood contamination. One or more species of bacteria were isolated from 55 (86%) of the handles, and included organisms such as enterococci, meticillin-susceptible Staphylococcus aureus, Klebsiella and acinetobacter. Cultures did not yield any anaerobes, fungi, meticillin-resistant Staphylococcus aureus, vancomycin-resistant enterococci or multiply-resistant Gram-negative bacilli. No occult blood contamination was demonstrated. Although the majority of organisms isolated were not pathogenic, their presence indicates the potential for transmission of pathogens from laryngoscope handles. Strategies to address contamination of handles include revision of procedures for disinfection and storage prior to use, introduction of disposable handles or sheaths, and re-design of handles to eliminate knurled surfaces and contact points.
Similar articles
-
Laryngoscope handles: a potential for infection.AANA J. 2000 Jun;68(3):233-6. AANA J. 2000. PMID: 11132011
-
Incidence of visible and occult blood on laryngoscope blades and handles.AANA J. 1997 Jun;65(3):241-6. AANA J. 1997. PMID: 9233093
-
Nosocomial contamination of laryngoscope handles: challenging current guidelines.Anesth Analg. 2009 Aug;109(2):479-83. doi: 10.1213/ane.0b013e3181ac1080. Anesth Analg. 2009. PMID: 19608822 Free PMC article.
-
Infection control practices of laryngoscope blades: a review of the literature.AANA J. 2012 Aug;80(4):274-8. AANA J. 2012. PMID: 23251996 Review.
-
Laryngoscope blades and handles as sources of cross-infection: an integrative review.J Hosp Infect. 2013 Apr;83(4):269-75. doi: 10.1016/j.jhin.2012.10.015. Epub 2013 Jan 16. J Hosp Infect. 2013. PMID: 23332194 Review.
Cited by
-
Prevention of contamination after endotracheal intubation using a dedicated sleeve.J Thorac Dis. 2023 Sep 28;15(9):4717-4724. doi: 10.21037/jtd-22-1510. Epub 2023 Aug 29. J Thorac Dis. 2023. PMID: 37868844 Free PMC article.
-
Comparison of efficacy and cost-effectiveness of 0.55% ortho-phthalaldehyde and 2% glutaraldehyde for disinfection of laryngoscopes: A prospective pilot study.Indian J Anaesth. 2017 Jun;61(6):490-493. doi: 10.4103/ija.IJA_22_17. Indian J Anaesth. 2017. PMID: 28655955 Free PMC article.
-
Surface contamination in the operating room: use of adenosine triphosphate monitoring.J Anesth. 2019 Feb;33(1):85-89. doi: 10.1007/s00540-018-2590-9. Epub 2018 Dec 11. J Anesth. 2019. PMID: 30539274
-
Pattern of anaesthetic equipment contamination and infection prevention in anaesthesia practice at university hospitals.Indian J Anaesth. 2018 Oct;62(10):786-792. doi: 10.4103/ija.IJA_41_18. Indian J Anaesth. 2018. PMID: 30443062 Free PMC article.
-
Bacterial Contamination and Disinfection Status of Laryngoscopes Stored in Emergency Crash Carts.J Prev Med Public Health. 2017;50(3):158-164. doi: 10.3961/jpmph.17.013. J Prev Med Public Health. 2017. PMID: 28605891 Free PMC article.
MeSH terms
LinkOut - more resources
Full Text Sources