. 2010 Feb 15;878(5-6):583-9.
doi: 10.1016/j.jchromb.2009.12.031. Epub 2010 Jan 4.

Quantitative determination of BAF312, a S1P-R modulator, in human urine by LC-MS/MS: prevention and recovery of lost analyte due to container surface adsorption

Affiliations

Quantitative determination of BAF312, a S1P-R modulator, in human urine by LC-MS/MS: prevention and recovery of lost analyte due to container surface adsorption

Wenkui Li et al. J Chromatogr B Analyt Technol Biomed Life Sci. .

Abstract

Analyte loss due to non-specific binding, especially container surface adsorption, is not uncommon in the quantitative analysis of urine samples. In developing a sensitive LC-MS/MS method for the determination of a drug candidate, BAF312, in human urine, a simple procedure was outlined for identification, confirmation and prevention of analyte non-specific binding to a container surface and to recover the 'non-specific loss' of an analyte, if no transfer has occurred to the original urine samples. Non-specific binding or container surface adsorption can be quickly identified by using freshly spiked urine calibration standards and pre-pooled QC samples during a LC-MS/MS feasibility run. The resulting low recovery of an analyte in urine samples can be prevented through the use of additives, such as the non-ionic surfactant Tween-80, CHAPS and others, to the container prior to urine sample collection. If the urine samples have not been transferred from the bulk container, the 'non-specific binding' of an analyte to the container surface can be reversed by the addition of a specified amount of CHAPS, Tween-80 or bovine serum albumin, followed by appropriate mixing. Among the above agents, Tween-80 is the most cost-effective. beta-cyclodextrin may be suitable in stabilizing the analyte of interest in urine via pre-treating the matrix with the agent. However, post-addition of beta-cyclodextrin to untreated urine samples does not recover the 'lost' analyte due to non-specific binding or container surface adsorption. In the case of BAF312, a dynamic range of 0.0200-20.0 ng/ml in human urine was validated with an overall accuracy and precision for QC sample results ranging from -3.2 to 5.1% (bias) and 3.9 to 10.2% (CV), respectively. Pre- and post-addition of 0.5% (v/v) Tween-80 to the container provided excellent overall analyte recovery and minimal MS signal suppression when a liquid-liquid extraction in combination with an isocratic LC separation was employed. The compound was stable in 0.5% Tween-80 treated human urine QC samples for at least 24 h at room temperature, after three freeze/thaw cycles with storage at < or =-60 degrees C and for at least 3 months when stored at < or =-60 degrees C. The current work could serve as a simple example in trouble shooting non-specific binding or container surface adsorption in quantitative analysis of urine samples.