Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
. 2010 Feb;63(2):349-55.
doi: 10.1002/mrm.22202.

Improved SNR in phase contrast velocimetry with five-point balanced flow encoding

Affiliations

Improved SNR in phase contrast velocimetry with five-point balanced flow encoding

Kevin M Johnson et al. Magn Reson Med. 2010 Feb.

Abstract

Phase contrast velocimetry can be utilized to measure complex flow for both quantitative and qualitative assessment of vascular hemodynamics. However, phase contrast requires that a maximum measurable velocity be set that balances noise and phase aliasing. To efficiently reduce noise in phase contrast images, several investigators have proposed extended velocity encoding schemes that use extra encodings to unwrap phase aliasing; however, existing techniques can lead to significant increases in echo and scan time, limiting their clinical benefits. In this work, we have developed a novel five-point velocity encoding scheme that efficiently reduces noise with minimal increases in scan and echo time. Investigations were performed in phantoms, demonstrating a 63% increase in velocity-to-noise ratio compared to standard four-point encoding schemes. Aortic velocity measurements were performed in healthy volunteers, showing similar velocity-to-noise ratio improvements. In those volunteers, it was also demonstrated that, without sacrificing accuracy, low-resolution images can be used for the fifth encoding point, reducing the scan time penalty from 25% down to less than 1%.

PubMed Disclaimer

Figures

Figure 1
Figure 1
Volume rendered aliasing spaces for 4-point balanced, 4-point referenced, and 5-point balanced encoding schemes viewed from 45° and -45° in the xy plane. Each shaded region has sphere rendered at the set Venc. In certain directions, all methods can measure velocities higher than the prescribed Venc, which is dependent on the encoding scheme. 4-point balanced encoding is particularly dependent on the velocity direction due to the rhombohedron shape, especially when compared to the cube and regular octahedron shapes of the 4-point referenced and 5-point balanced techniques, respectively.
Figure 2
Figure 2
Magnitude image showing phantom rotation (a) and corresponding left to right velocity images. 4-point balanced images (b) have the same first moment magnitude as the proposed 5-point scheme (c), while 4-point referenced images (d) have the same Venc as the 5-point exam. 4-point balanced images show significant aliasing, which is not evident in 5-point images. Visually, 5-point velocity images show improved VNR over 4-point referenced images.
Figure 3
Figure 3
Source velocity images from 4-point referenced, 4-point balanced, and 5-point acquisitions. Images from the 5-point exam show reduced noise compared to the equal Venc exam (Referenced) and less velocity aliasing than the same first moment exam (Balanced).
Figure 4
Figure 4
In-plane vector plots during systole (t=143ms) and during peak flow reversal (t=469ms) for 5-point balanced encoding and 4-point referenced encoding. Systole images show the location of vector plots on the magnitude images. During systole, both encoding schemes produce visually similar images. During diastole, where there velocities are substantially lower, 5-point encoding shows substantially reduced noise for better visualization.
Figure 5
Figure 5
Source S/I velocity images for a single volunteer (a) and measured aliasing (b) as a function of resolution for accelerated 5-point encoding. On source images, arrows point to locations of uncorrected velocity aliasing, while images are labeled with the percent increase in scan time. The no aliasing area is indicated on the measured aliasing and is bounded by a 23.8mm spatial resolution and 331ms temporal resolution. Error is significantly more dependent on spatial resolution than temporal resolution in this case.

Similar articles

Cited by

References

    1. Markl M, Harloff A, Bley TA, Zaitsev M, Jung B, Weigang E, Langer M, Hennig J, Frydrychowicz A. Time-resolved 3D MR velocity mapping at 3T: improved navigator-gated assessment of vascular anatomy and blood flow. J Magn Reson Imaging. 2007;25(4):824–831. - PubMed
    1. Kvitting JP, Ebbers T, Wigstrom L, Engvall J, Olin CL, Bolger AF. Flow patterns in the aortic root and the aorta studied with time-resolved, 3-dimensional, phase-contrast magnetic resonance imaging: implications for aortic valve-sparing surgery. J Thorac Cardiovasc Surg. 2004;127(6):1602–1607. - PubMed
    1. Wigstrom L, Ebbers T, Fyrenius A, Karlsson M, Engvall J, Wranne B, Bolger AF. Particle trace visualization of intracardiac flow using time-resolved 3D phase contrast MRI. Magn Reson Med. 1999;41(4):793–799. - PubMed
    1. Stalder AF, Russe MF, Frydrychowicz A, Bock J, Hennig J, Markl M. Quantitative 2D and 3D phase contrast MRI: Optimized analysis of blood flow and vessel wall parameters. Magn Reson Med. 2008;60(5):1218–1231. - PubMed
    1. Wu SP, Ringgaard S, Pedersen EM. Three-dimensional phase contrast velocity mapping acquisition improves wall shear stress estimation in vivo. Magn Reson Imaging. 2004;22(3):345–351. - PubMed

Publication types

MeSH terms