Design and evaluation of a 32-channel phased-array coil for lung imaging with hyperpolarized 3-helium
- PMID: 20099333
- DOI: 10.1002/mrm.22265
Design and evaluation of a 32-channel phased-array coil for lung imaging with hyperpolarized 3-helium
Abstract
Imaging with hyperpolarized 3-helium is becoming an increasingly important technique for MRI diagnostics of the lung but is hampered by long breath holds (>20 sec), which are not always applicable in patients with severe lung disease like chronic obstructive pulmonary disease (COPD) or alpha-1-anti-trypsin deficiency. Additionally, oxygen-induced depolarization decay during the long breath holds complicates interpretation of functional data such as apparent diffusion coefficients. To address these issues, we describe and validate a 1.5-T, 32-channel array coil for accelerated (3)He lung imaging and demonstrate its ability to speed up imaging (3)He. A signal-to-noise ratio increase of up to a factor of 17 was observed compared to a conventional double-resonant birdcage for unaccelerated imaging, potentially allowing increased image resolution or decreased gas production requirements. Accelerated imaging of the whole lung with one-dimensional and two-dimensional acceleration factors of 4 and 4 x 2, respectively, was achieved while still retaining excellent image quality. Finally, the potential of highly parallel detection in lung imaging is demonstrated with high-resolution morphologic and functional images.
Similar articles
-
A flexible 32-channel receive array combined with a homogeneous transmit coil for human lung imaging with hyperpolarized 3He at 1.5 T.Magn Reson Med. 2011 Dec;66(6):1788-97. doi: 10.1002/mrm.22962. Epub 2011 May 13. Magn Reson Med. 2011. PMID: 21574180
-
Asymmetric quadrature split birdcage coil for hyperpolarized 3He lung MRI at 1.5T.Magn Reson Med. 2008 Aug;60(2):431-8. doi: 10.1002/mrm.21664. Magn Reson Med. 2008. PMID: 18666111
-
Advantages of parallel imaging in conjunction with hyperpolarized helium--a new approach to MRI of the lung.Magn Reson Med. 2006 May;55(5):1132-41. doi: 10.1002/mrm.20855. Magn Reson Med. 2006. PMID: 16586457
-
[Functional imaging of the lung using a gaseous contrast agent: (3)helium-magnetic resonance imaging].Rofo. 2005 May;177(5):660-9. doi: 10.1055/s-2005-858063. Rofo. 2005. PMID: 15871081 Review. German.
-
Functional MRI of the lung using hyperpolarized 3-helium gas.J Magn Reson Imaging. 2004 Oct;20(4):540-54. doi: 10.1002/jmri.20154. J Magn Reson Imaging. 2004. PMID: 15390146 Review.
Cited by
-
Hyperpolarized 129Xe MRI of the human lung.J Magn Reson Imaging. 2013 Feb;37(2):313-31. doi: 10.1002/jmri.23844. J Magn Reson Imaging. 2013. PMID: 23355432 Free PMC article. Review.
-
In vivo lung morphometry with accelerated hyperpolarized (3) He diffusion MRI: a preliminary study.Magn Reson Med. 2015 Apr;73(4):1609-14. doi: 10.1002/mrm.25284. Epub 2014 May 5. Magn Reson Med. 2015. PMID: 24799044 Free PMC article.
-
In vivo methods and applications of xenon-129 magnetic resonance.Prog Nucl Magn Reson Spectrosc. 2021 Feb;122:42-62. doi: 10.1016/j.pnmrs.2020.11.002. Epub 2020 Dec 9. Prog Nucl Magn Reson Spectrosc. 2021. PMID: 33632417 Free PMC article. Review.
-
Assessment of a novel 32-channel phased array for cardiovascular hybrid PET/MRI imaging: MRI performance.Eur J Hybrid Imaging. 2019;3(1):13. doi: 10.1186/s41824-019-0061-7. Epub 2019 Aug 22. Eur J Hybrid Imaging. 2019. PMID: 33283144 Free PMC article.
-
A 16-Channel 13C Array Coil for Magnetic Resonance Spectroscopy of the Breast at 7T.IEEE Trans Biomed Eng. 2021 Jun;68(6):2036-2046. doi: 10.1109/TBME.2021.3063061. Epub 2021 May 21. IEEE Trans Biomed Eng. 2021. PMID: 33651680 Free PMC article.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Medical