Device-related infection among patients with pacemakers and implantable defibrillators: incidence, risk factors, and consequences
- PMID: 20102431
- DOI: 10.1111/j.1540-8167.2009.01690.x
Device-related infection among patients with pacemakers and implantable defibrillators: incidence, risk factors, and consequences
Abstract
Background: Device-related infection is a major limitation of device therapy for cardiac arrhythmia.
Methods: The authors analyzed the incidence and risk factors for cardiac device infection (CDI) among consecutive patients implanted with pacemaker (PM) or implantable cardioverter defibrillator (ICD) (including cardiac resynchronization therapy devices) at a tertiary health center in Hamilton, Ontario, Canada. Most patients with device-related infections were identified by an internal infection control system that reports any positive wound and blood cultures following surgery, between 2005 and the present. A retrospective review of patient records was also performed for all patients who received an ICD or PM between July 1, 2003 and March 20, 2007.
Results: A total of 24 infections were identified among 2,417 patients having device surgery (1%). Fifteen of these infections (60%) were diagnosed within 90 days of the last surgical procedure. Univariate analysis showed that patients presenting with CDI were more likely to have had a device replacement, rather than a new implant, had more complex devices (dual/triple chamber vs single), and were more likely to have had a prior lead dislodgement. Multivariate analysis found device replacement (P = 0.02) and cardiac resynchronization therapy (CRT)/dual-chamber devices (P = 0.048) to be independent predictors of infection. One patient developed septic pulmonary emboli after having laser-assisted lead extraction. No patient died and 22 patients received a new device.
Conclusion: CDI occurs in about 1% of cases in high volume facilities. Pulse generator replacement surgery and dual- or triple-chamber device implantation were associated with a significantly increased risk of infection.
Similar articles
-
Management and outcome of permanent pacemaker and implantable cardioverter-defibrillator infections.J Am Coll Cardiol. 2007 May 8;49(18):1851-9. doi: 10.1016/j.jacc.2007.01.072. Epub 2007 Apr 23. J Am Coll Cardiol. 2007. PMID: 17481444
-
Risk factors related to infections of implanted pacemakers and cardioverter-defibrillators: results of a large prospective study.Circulation. 2007 Sep 18;116(12):1349-55. doi: 10.1161/CIRCULATIONAHA.106.678664. Epub 2007 Aug 27. Circulation. 2007. PMID: 17724263
-
Efficacy of antibiotic prophylaxis before the implantation of pacemakers and cardioverter-defibrillators: results of a large, prospective, randomized, double-blinded, placebo-controlled trial.Circ Arrhythm Electrophysiol. 2009 Feb;2(1):29-34. doi: 10.1161/CIRCEP.108.795906. Epub 2009 Feb 10. Circ Arrhythm Electrophysiol. 2009. PMID: 19808441 Clinical Trial.
-
[Pacemaker and implantable cardioverter defibrillator infections].Med Mal Infect. 2010 Aug;40(8):429-39. doi: 10.1016/j.medmal.2009.11.005. Epub 2009 Dec 30. Med Mal Infect. 2010. PMID: 20044226 Review. French.
-
[Non-valvular cardiac devices endocarditis].Ann Cardiol Angeiol (Paris). 2008 Apr;57(2):81-7. doi: 10.1016/j.ancard.2008.02.005. Epub 2008 Mar 18. Ann Cardiol Angeiol (Paris). 2008. PMID: 18402924 Review. French.
Cited by
-
Outcomes after endocarditis or device infection in patients with left ventricular epicardial leads versus coronary sinus leads.J Interv Card Electrophysiol. 2014 Apr;39(3):267-71. doi: 10.1007/s10840-014-9880-4. Epub 2014 Feb 23. J Interv Card Electrophysiol. 2014. PMID: 24562754
-
Making decisions about implantable cardioverter-defibrillators from implantation to end of life: an integrative review of patients' perspectives.Patient. 2014;7(3):243-60. doi: 10.1007/s40271-014-0055-2. Patient. 2014. PMID: 24668214 Review.
-
A case of defibrillator-associated infective endocarditis due to Campylobacter fetus.Folia Microbiol (Praha). 2016 Nov;61(6):529-532. doi: 10.1007/s12223-016-0466-7. Epub 2016 Jul 29. Folia Microbiol (Praha). 2016. PMID: 27473849
-
Conservative therapy for the management of cardiac implantable electronic device infection.J Arrhythm. 2016 Aug;32(4):293-6. doi: 10.1016/j.joa.2015.09.012. Epub 2015 Nov 19. J Arrhythm. 2016. PMID: 27588152 Free PMC article. Review.
-
Early, Delayed and Late Cardiac Implantable Electronic Device Infections: Do the Timing of Onset and Pathogens Matter?J Clin Med. 2022 Jul 6;11(14):3929. doi: 10.3390/jcm11143929. J Clin Med. 2022. PMID: 35887692 Free PMC article.
Publication types
MeSH terms
Substances
Grants and funding
LinkOut - more resources
Full Text Sources
Medical
Research Materials