Accuracy of a new ring-opening metathesis elastomeric dental impression material with spray and immersion disinfection
- PMID: 20105678
- DOI: 10.1016/S0022-3913(09)60210-9
Accuracy of a new ring-opening metathesis elastomeric dental impression material with spray and immersion disinfection
Abstract
Statement of problem: A new elastomeric impression material has been formulated with a ring-opening metathesis chemistry. In addition to other properties of clinical significance, the impression accuracy must be confirmed.
Purpose: The purpose of this study was to compare the accuracy of the new elastomeric impression material with vinyl polysiloxane and polyether following both spray and immersion disinfection.
Material and methods: Impressions of a modified dentoform with a stainless steel crown preparation in the lower right quadrant were made, and type IV gypsum working casts and dies were formed. Anteroposterior (AP), cross-arch (CA), buccolingual (BL), mesiodistal (MD), occlusogingivobuccal (OGB), and occlusogingivolingual (OGL) dimensions were measured using a microscope. Working cast and die dimensions were compared to those of the master model. The impression materials were a newly formulated, ring-opening metathesis-polymerization impression material (ROMP Cartridge Tray and ROMP Volume Wash), vinyl polysiloxane (VPS, Aquasil Ultra Monophase/LV), and a polyether (PE, Impregum Penta Soft/Permadyne Garant L). Fifteen impressions with each material were made, of which 5 were disinfected by spray for 10 minutes (CaviCide), 5 were disinfected by immersion for 90 minutes (ProCide D), and 5 were not disinfected. There were significant cross-product interactions with a 2-way ANOVA, so a 1-way ANOVA and Dunnett's T3 multiple comparison test were used to compare the dimensional changes of the 3 impression materials, by disinfection status and for each location (alpha=.05).
Results: For ROMP, there were no significant differences from the master, for any dimension, when comparing the control and 2 disinfectant conditions. No significant differences were detected among the 3 impression materials for CA, BL, and MD. The working die dimensions of OGB and OGL for VPS with immersion disinfection were significantly shorter than with PE and ROMP (P<.05). Overall, the AP dimension was more accurate than CA, and the BL of working dies was 0.040 mm greater in diameter than MD.
Conclusions: The accuracy of gypsum working casts and working dies from the new and 2 existing types of impression material were similar, for both spray and immersion disinfection. Judicious application of a die spacer can compensate for the small differences observed. VPS may require additional laboratory accommodation to compensate for a shorter working die.
Copyright 2010 The Editorial Council of the Journal of Prosthetic Dentistry. Published by Mosby, Inc. All rights reserved.
Similar articles
-
Accuracy of newly formulated fast-setting elastomeric impression materials.J Prosthet Dent. 2005 Jun;93(6):530-9. doi: 10.1016/j.prosdent.2005.03.007. J Prosthet Dent. 2005. PMID: 15942613
-
Accuracy of the newly formulated vinyl siloxanether elastomeric impression material.J Prosthet Dent. 2010 Apr;103(4):228-39. doi: 10.1016/S0022-3913(10)60035-2. J Prosthet Dent. 2010. PMID: 20362766
-
Accuracy of a reformulated fast-set vinyl polysiloxane impression material using dual-arch trays.J Prosthet Dent. 2009 May;101(5):332-41. doi: 10.1016/S0022-3913(09)60066-4. J Prosthet Dent. 2009. PMID: 19410067
-
Accuracy and stability of impression materials subjected to chemical disinfection - a literature review.J Oral Rehabil. 2008 Apr;35(4):291-9. doi: 10.1111/j.1365-2842.2007.01771.x. J Oral Rehabil. 2008. PMID: 18321265 Review.
-
Dimensional stability of elastomeric impression materials: a critical review of the literature.Eur J Prosthodont Restor Dent. 2011 Dec;19(4):163-6. Eur J Prosthodont Restor Dent. 2011. PMID: 22645802 Review.
Cited by
-
Dimensional stability of two impression materials after a 6-month storage period.Acta Biomater Odontol Scand. 2017 Nov 14;3(1):84-91. doi: 10.1080/23337931.2017.1401933. eCollection 2017. Acta Biomater Odontol Scand. 2017. PMID: 29242816 Free PMC article.
-
A simple and effective method for addition silicone impression disinfection.J Adv Prosthodont. 2019 Jun;11(3):155-161. doi: 10.4047/jap.2019.11.3.155. Epub 2019 Jun 26. J Adv Prosthodont. 2019. PMID: 31297174 Free PMC article.
-
Comparative evaluation of the effect of electrolyzed oxidizing water on surface detail reproduction, dimensional stability and Surface texture of poly vinyl siloxane impressions.J Indian Prosthodont Soc. 2019 Jan-Mar;19(1):33-41. doi: 10.4103/jips.jips_72_18. J Indian Prosthodont Soc. 2019. PMID: 30745752 Free PMC article.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Research Materials
Miscellaneous