A protocol for patients with cardiovascular implantable devices undergoing magnetic resonance imaging (MRI): should defibrillation threshold testing be performed post-(MRI)
- PMID: 20111895
- DOI: 10.1007/s10840-009-9463-y
A protocol for patients with cardiovascular implantable devices undergoing magnetic resonance imaging (MRI): should defibrillation threshold testing be performed post-(MRI)
Abstract
Purpose: Magnetic resonance imaging (MRI) in patients with Cardiovascular Implantable Electronic Devices (CIED) has not been approved by the Food and Drug Administration. Recent data suggests MRI as a relative rather than absolute contraindication in CIED patients. Recently, the American Heart Association has recommended defibrillation threshold testing (DFTT) in implantable cardioverter defibrillator (ICD) patients undergoing MRI. We evaluated the feasibility and safety of a protocol for MRI in CIED patients, incorporating the new recommendations on DFTT.
Methods: Consecutive patients with CIED undergoing MRI were included. The protocol consisted of continuous monitoring during imaging, device interrogation pre- and post-MRI, reprogramming of the pacemaker to an asynchronous mode in pacemaker-dependent (PMD) patients and a non-tracking/sensing mode for non-PMD patients. All tachyarrhythmia therapies were disabled. Devices were interrogated for lead impedance, battery life, pacing, and sensing thresholds. All patients with ICD underwent DFTT/defibrillator safety margin testing (DSMT) post-MRI.
Results: A total of 92 MRI's at 1.5 Tesla were performed in 38 patients. A total of 13 PMD patients, ten ICD patients, four cardiac resynchronization therapy with defibrillator (CRT-D) patients, and 11 non-PMD patients were scanned from four major manufacturers. No device circuitry damage, programming alterations, inappropriate shocks, failure to pace, or changes in sensing, pacing, or defibrillator thresholds were found on single or multiple MRI sessions.
Conclusions: Our protocol for MRI in CIED patients appears safe, feasible, and reproducible. This is irrespective of the type of CIED, pacemaker dependancy or multiple 24-h scanning sessions. Our protocol addresses early detection of potential complications and establishes a response system for potential device-related complications. Our observation suggests that routine DFTT/DSMT post-MRI may not be necessary.
Comment in
-
Guidelines and the growing service burden.J Interv Card Electrophysiol. 2010 Aug;28(2):83-5. doi: 10.1007/s10840-010-9480-x. Epub 2010 Apr 13. J Interv Card Electrophysiol. 2010. PMID: 20386972 No abstract available.
Similar articles
-
Clinical utility and safety of a protocol for noncardiac and cardiac magnetic resonance imaging of patients with permanent pacemakers and implantable-cardioverter defibrillators at 1.5 tesla.Circulation. 2006 Sep 19;114(12):1277-84. doi: 10.1161/CIRCULATIONAHA.105.607655. Epub 2006 Sep 11. Circulation. 2006. PMID: 16966586 Free PMC article. Clinical Trial.
-
The Safety of Cardiac and Thoracic Magnetic Resonance Imaging in Patients with Cardiac Implantable Electronic Devices.Acad Radiol. 2016 Dec;23(12):1498-1505. doi: 10.1016/j.acra.2016.08.016. Epub 2016 Oct 4. Acad Radiol. 2016. PMID: 27717762
-
Magnetic resonance imaging in patients with cardiac implantable electronic devices: a single-center prospective study.J Interv Card Electrophysiol. 2017 Oct;50(1):95-104. doi: 10.1007/s10840-017-0262-6. Epub 2017 Jul 22. J Interv Card Electrophysiol. 2017. PMID: 28733766
-
[Cardiac implantable electronic devices (CIED) not an absolute contraindication to MRI].Lakartidningen. 2017 Mar 21;114:ECLR. Lakartidningen. 2017. PMID: 28350417 Review. Swedish.
-
Magnetic resonance imaging safety in nonconditional pacemaker and defibrillator recipients: A meta-analysis and systematic review.Heart Rhythm. 2018 Jul;15(7):1001-1008. doi: 10.1016/j.hrthm.2018.02.019. Epub 2018 Feb 17. Heart Rhythm. 2018. PMID: 29458192
Cited by
-
[Magnetic resonance imaging and implantable cardiac devices. Current status and future perspectives of MR-compatible systems].Herz. 2012 Mar;37(2):136-45. doi: 10.1007/s00059-012-3588-y. Herz. 2012. PMID: 22398815 German.
-
Guidelines and the growing service burden.J Interv Card Electrophysiol. 2010 Aug;28(2):83-5. doi: 10.1007/s10840-010-9480-x. Epub 2010 Apr 13. J Interv Card Electrophysiol. 2010. PMID: 20386972 No abstract available.
-
Guideline for Cardiovascular Magnetic Resonance Imaging from the Korean Society of Cardiovascular Imaging-Part 1: Standardized Protocol.Korean J Radiol. 2019 Sep;20(9):1313-1333. doi: 10.3348/kjr.2019.0398. Korean J Radiol. 2019. PMID: 31464111 Free PMC article.
-
Echocardiographic parameters to predict inadequate defibrillation safety margin in patients receiving implantable cardioverter defibrillators for primary prevention.J Interv Card Electrophysiol. 2013 Jun;37(1):79-85. doi: 10.1007/s10840-012-9764-4. Epub 2013 Jan 20. J Interv Card Electrophysiol. 2013. PMID: 23334901 Clinical Trial.
-
Issues with radiofrequency heating in MRI.J Appl Clin Med Phys. 2014 Sep 8;15(5):5064. doi: 10.1120/jacmp.v15i5.5064. J Appl Clin Med Phys. 2014. PMID: 25207583 Free PMC article. No abstract available.
References
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical
Research Materials