Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
. 1991 Jan 25;51(1):19-32.

[Fast MR imaging of liver metastasis using FLASH and FISP--optimal sequences for T1- and T2*-weighted images]

[Article in Japanese]
Affiliations
  • PMID: 2011477

[Fast MR imaging of liver metastasis using FLASH and FISP--optimal sequences for T1- and T2*-weighted images]

[Article in Japanese]
S Otake et al. Nihon Igaku Hoshasen Gakkai Zasshi. .

Abstract

This paper deals with a study to obtain the optimal sequence of gradient echo (GE) for T1- and T2*-weighted images similar to T1- and T2-weighted images of spin echo (SE). Two GE sequences, fast low angle shot (FLASH) and fast imaging with steady-state precession (FISP), were performed in 15 cases of liver metastasis in various combination of flip angle (FA), repetition time (TR), and echo time (TE). The optimal combinations were summarized as follows: 1) T1-weighted FLASH image with FA of 40 degrees, TR of 22 msec and TE of 10 msec, 2) T1-weighted FISP image with FA of 70 degrees, TR of 100 msec, TE of 10 msec, 3) both T2*-weighted FLASH and FISP images with FA of 10 degrees, TR of 100 msec and TE of 30 msec. Not only to provide the adequate T1- and T2*-weighted images but also to enable breath-holding MR imaging, GE sequences can optionally take place SE in cases of deteriorated images caused by moving artifacts. Other applications support the re-examination and further detailing when required, conveniently rather in short time.

PubMed Disclaimer

Similar articles

Publication types

LinkOut - more resources