Radiation dose reduction with chest computed tomography using adaptive statistical iterative reconstruction technique: initial experience
Abstract
Purpose: To assess radiation dose reduction and image quality for weight-based chest computed tomographic (CT) examination results reconstructed using adaptive statistical iterative reconstruction (ASIR) technique.
Materials and methods: With local ethical committee approval, weight-adjusted chest CT examinations were performed using ASIR in 98 patients and filtered backprojection (FBP) in 54 weight-matched patients on a 64-slice multidetector CT. Patients were categorized into 3 groups: 60 kg or less (n = 32), 61 to 90 kg (n = 77), and 91 kg or more (n = 43) for weight-based adjustment of noise indices for automatic exposure control (Auto mA; GE Healthcare, Waukesha, Wis). Remaining scan parameters were held constant at 0.984:1 pitch, 120 kilovolts (peak), 40-mm table feed per rotation, and 2.5-mm section thickness. Patients' weight, scanning parameters, and CT dose index volume were recorded. Effective doses (EDs) were estimated. Image noise was measured in the descending thoracic aorta at the level of the carina. Data were analyzed using analysis of variance.
Results: Compared with FBP, ASIR was associated with an overall mean (SD) decrease of 27.6% in ED (ASIR, 8.8 [2.3] mSv; FBP, 12.2 [2.1] mSv; P < 0.0001). With the use of ASIR, the ED values were 6.5 (1.8) mSv (28.8% decrease), 7.3 (1.6) mSv (27.3% decrease), and 12.8 (2.3) mSv (26.8% decrease) for the weight groups of 60 kg or less, 61 to 90 kg, and 91 kg or more, respectively, compared with 9.2 (2.3) mSv, 10.0 (2.0) mSv, and 17.4 (2.1) mSv with FBP (P < 0.0001). Despite dose reduction, there was less noise with ASIR (12.6 [2.9] mSv) than with FBP (16.6 [6.2] mSv; P < 0.0001).
Conclusions: Adaptive statistical iterative reconstruction helps reduce chest CT radiation dose and improve image quality compared with the conventionally used FBP image reconstruction.
Similar articles
-
Reducing abdominal CT radiation dose with adaptive statistical iterative reconstruction technique.Invest Radiol. 2010 Apr;45(4):202-10. doi: 10.1097/RLI.ob013e3181dzfeec. Invest Radiol. 2010. PMID: 20177389
-
Comparison of hybrid and pure iterative reconstruction techniques with conventional filtered back projection: dose reduction potential in the abdomen.J Comput Assist Tomogr. 2012 May-Jun;36(3):347-53. doi: 10.1097/RCT.0b013e31824e639e. J Comput Assist Tomogr. 2012. PMID: 22592622
-
Radiation dose reduction with hybrid iterative reconstruction for pediatric CT.Radiology. 2012 May;263(2):537-46. doi: 10.1148/radiol.12110268. Radiology. 2012. PMID: 22517962
-
Iterative reconstruction of pulmonary MDCT angiography: effects on image quality, effective dose and estimated organ dose to the breast.Biomed Pap Med Fac Univ Palacky Olomouc Czech Repub. 2014 Jun;158(2):259-64. doi: 10.5507/bp.2013.059. Epub 2013 Sep 12. Biomed Pap Med Fac Univ Palacky Olomouc Czech Repub. 2014. PMID: 24026145 Review.
-
Radiation dose reduction in chest CT: a review.AJR Am J Roentgenol. 2008 Feb;190(2):335-43. doi: 10.2214/AJR.07.2556. AJR Am J Roentgenol. 2008. PMID: 18212218 Review.
Cited by
-
Radiation dose and image quality in pediatric chest CT: effects of iterative reconstruction in normal weight and overweight children.Pediatr Radiol. 2015 Mar;45(3):337-44. doi: 10.1007/s00247-014-3176-9. Epub 2014 Sep 26. Pediatr Radiol. 2015. PMID: 25256153
-
Feasible Dose Reduction in Routine Chest Computed Tomography Maintaining Constant Image Quality Using the Last Three Scanner Generations: From Filtered Back Projection to Sinogram-affirmed Iterative Reconstruction and Impact of the Novel Fully Integrated Detector Design Minimizing Electronic Noise.J Clin Imaging Sci. 2014 Jul 31;4:38. doi: 10.4103/2156-7514.137826. eCollection 2014. J Clin Imaging Sci. 2014. PMID: 25161807 Free PMC article.
-
New adaptive statistical iterative reconstruction ASiR-V: Assessment of noise performance in comparison to ASiR.J Appl Clin Med Phys. 2018 Mar;19(2):275-286. doi: 10.1002/acm2.12253. Epub 2018 Jan 24. J Appl Clin Med Phys. 2018. PMID: 29363260 Free PMC article.
-
Effects of adaptive statistical iterative reconstruction on radiation dose reduction and diagnostic accuracy of pediatric abdominal CT.Pediatr Radiol. 2014 Dec;44(12):1541-7. doi: 10.1007/s00247-014-3058-1. Epub 2014 Jul 8. Pediatr Radiol. 2014. PMID: 25001398
-
Optimisation of low-dose CT with adaptive statistical iterative reconstruction in total body examination.Radiol Med. 2012 Dec;117(8):1333-46. doi: 10.1007/s11547-012-0897-3. Epub 2012 Oct 22. Radiol Med. 2012. PMID: 23090256 English, Italian.
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical
Miscellaneous