Using 100- instead of 120-kVp computed tomography to diagnose pulmonary embolism almost halves the radiation dose with preserved diagnostic quality
- PMID: 20121673
- DOI: 10.3109/02841850903505222
Using 100- instead of 120-kVp computed tomography to diagnose pulmonary embolism almost halves the radiation dose with preserved diagnostic quality
Abstract
Background: Concern has been raised regarding the mounting collective radiation doses from computed tomography (CT), increasing the risk of radiation-induced cancers in exposed populations.
Purpose: To compare radiation dose and image quality in a chest phantom and in patients for the diagnosis of pulmonary embolism (PE) at 100 and 120 peak kilovoltage (kVp) using 16-multichannel detector computed tomography (MDCT).
Material and methods: A 20-ml syringe containing 12 mg I/ml was scanned in a chest phantom at 100/120 kVp and 25 milliampere seconds (mAs). Consecutive patients underwent 100 kVp (n = 50) and 120 kVp (n = 50) 16-MDCT using a "quality reference" effective mAs of 100, 300 mg I/kg, and a 12-s injection duration. Attenuation (CT number), image noise (1 standard deviation), and contrast-to-noise ratio (CNR; fresh clot = 70 HU) of the contrast medium syringe and pulmonary arteries were evaluated on 3-mm-thick slices. Subjective image quality was assessed. Computed tomography dose index (CTDI(vol)) and dose-length product (DLP) were presented by the CT software, and effective dose was estimated.
Results: Mean values in the chest phantom and patients changed as follows when X-ray tube potential decreased from 120 to 100 kVp: attenuation +23% and +40%, noise +38% and +48%, CNR -6% and 0%, and CTDI(vol) -38% and -40%, respectively. Mean DLP and effective dose in the patients decreased by 42% and 45%, respectively. Subjective image quality was excellent or adequate in 49/48 patients at 100/120 kVp. No patient with a negative CT had any thromboembolism diagnosed during 3-month follow-up.
Conclusion: By reducing X-ray tube potential from 120 to 100 kVp, while keeping all other scanning parameters unchanged, the radiation dose to the patient may be almost halved without deterioration of diagnostic quality, which may be of particular benefit in young individuals.
Similar articles
-
Low-dose radiation with 80-kVp computed tomography to diagnose pulmonary embolism: a feasibility study.Acta Radiol. 2012 Nov 1;53(9):1004-13. doi: 10.1258/ar.2012.120327. Epub 2012 Aug 31. Acta Radiol. 2012. PMID: 22940858
-
Minimizing contrast medium doses to diagnose pulmonary embolism with 80-kVp multidetector computed tomography in azotemic patients.Acta Radiol. 2009 Mar;50(2):181-93. doi: 10.1080/02841850802657269. Acta Radiol. 2009. PMID: 19169917
-
Dual-energy, standard and low-kVp contrast-enhanced CT-cholangiography: a comparative analysis of image quality and radiation exposure.Eur J Radiol. 2012 Jul;81(7):1405-12. doi: 10.1016/j.ejrad.2011.03.030. Epub 2011 Apr 2. Eur J Radiol. 2012. PMID: 21458939
-
Pulmonary CT angiography as first-line imaging for PE: image quality and radiation dose considerations.AJR Am J Roentgenol. 2013 Mar;200(3):522-8. doi: 10.2214/AJR.12.9928. AJR Am J Roentgenol. 2013. PMID: 23436840 Review.
-
Dual-energy computed tomography imaging of the aorta.J Thorac Imaging. 2010 Nov;25(4):289-300. doi: 10.1097/RTI.0b013e3181dc2b4c. J Thorac Imaging. 2010. PMID: 21042067 Review.
Cited by
-
Incidental findings detection using low tube potential for CT pulmonary angiography.Int J Cardiovasc Imaging. 2014 Dec;30(8):1579-88. doi: 10.1007/s10554-014-0484-y. Epub 2014 Aug 1. Int J Cardiovasc Imaging. 2014. PMID: 25082644
-
Image quality of spectral brain computed tomography angiography using halved dose of iodine contrast medium.Neuroradiology. 2023 Sep;65(9):1333-1342. doi: 10.1007/s00234-023-03190-1. Epub 2023 Jul 15. Neuroradiology. 2023. PMID: 37452885 Free PMC article.
-
Carotid CTA: radiation exposure and image quality with the use of attenuation-based, automated kilovolt selection.AJNR Am J Neuroradiol. 2014 Feb;35(2):237-41. doi: 10.3174/ajnr.A3659. Epub 2013 Aug 1. AJNR Am J Neuroradiol. 2014. PMID: 23907241 Free PMC article.
-
Use of Model-Based Iterative Reconstruction (MBIR) in reduced-dose CT for routine follow-up of patients with malignant lymphoma: dose savings, image quality and phantom study.Eur Radiol. 2015 Aug;25(8):2362-70. doi: 10.1007/s00330-015-3656-9. Epub 2015 Mar 8. Eur Radiol. 2015. PMID: 25749813
-
CT pulmonary angiography: simultaneous low-pitch dual-source acquisition mode with 70 kVp and 40 ml of contrast medium and comparison with high-pitch spiral dual-source acquisition with automated tube potential selection.Br J Radiol. 2016 Jun;89(1062):20151059. doi: 10.1259/bjr.20151059. Epub 2016 Mar 23. Br J Radiol. 2016. PMID: 27007972 Free PMC article.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Medical