Discrepancies between two automated immunoassay systems in determining hepatitis B virus markers in serum samples with concomitant presence of antigens and antibodies
- PMID: 20124330
Discrepancies between two automated immunoassay systems in determining hepatitis B virus markers in serum samples with concomitant presence of antigens and antibodies
Abstract
This study compared the assay results of HBV markers obtained with two automated immunoassay systems in serum samples with concomitant HBV antigens and antibodies. Eighty-nine samples of concomitant HBsAg and anti-HBs and 74 samples of concomitant HBeAg and anti-HBe were analyzed by the Abbott Architect i2000 and the Roche Elecsys 2010 immunoassay systems. The overall concordance rates of the Architect and Elecsys results were 78.6% for samples with concomitant HBsAg and anti-HBs and 77% for samples with concomitant HBeAg and anti-HBe. The data show that substantial differences exist between the results obtained by these two systems, especially in determining anti-HBs and HBeAg at low concentrations.
Similar articles
-
Comparison of the results for three automated immunoassay systems in determining serum HBV markers.Clin Chim Acta. 2006 Oct;372(1-2):129-33. doi: 10.1016/j.cca.2006.03.032. Epub 2006 May 19. Clin Chim Acta. 2006. PMID: 16713592
-
Comparison of the Abbott Architect i2000 assay, the Roche Modular Analytics E170 assay, and an immunoradiometric assay for serum hepatitis B virus markers.Ann Clin Lab Sci. 2007 Summer;37(3):256-9. Ann Clin Lab Sci. 2007. PMID: 17709690
-
Performance evaluation of the ADVIA Centaur anti-HBe and HBeAg assays.J Clin Virol. 2008 Oct;43(2):169-75. doi: 10.1016/j.jcv.2008.05.008. Epub 2008 Jul 16. J Clin Virol. 2008. PMID: 18635392
-
[Diagnostic tests: Serological markers of hepatitis B virus].Nihon Rinsho. 2005 Jul;63 Suppl 7:404-7. Nihon Rinsho. 2005. PMID: 16111286 Review. Japanese. No abstract available.
-
Seroprevalence of markers for hepatitis B viral infection.Int J Infect Dis. 2011 Feb;15(2):e78-121. doi: 10.1016/j.ijid.2010.09.005. Epub 2010 Dec 4. Int J Infect Dis. 2011. PMID: 21130675 Review.
Cited by
-
The Performance of the Abbott i2000 for Measuring Serum Markers of Infectious Diseases.J Clin Lab Anal. 2017 Jan;31(1):e22015. doi: 10.1002/jcla.22015. Epub 2016 Jun 15. J Clin Lab Anal. 2017. PMID: 27302692 Free PMC article.
-
Cost-Effective In-House Neutralization Assay for the Confirmation of HBeAg.J Clin Lab Anal. 2016 Nov;30(6):1146-1149. doi: 10.1002/jcla.21995. Epub 2016 May 21. J Clin Lab Anal. 2016. PMID: 27207380 Free PMC article.
-
Comparison of Autolumo A2000 Plus and Architect i2000 for detection of hepatitis B virus serological markers.Heliyon. 2024 Jun 8;10(14):e32698. doi: 10.1016/j.heliyon.2024.e32698. eCollection 2024 Jul 30. Heliyon. 2024. PMID: 39113996 Free PMC article.
-
A simple and rapid capillary chemiluminescence immunoassay for quantitatively detecting human serum HBsAg.Eur J Clin Microbiol Infect Dis. 2013 Dec;32(12):1557-64. doi: 10.1007/s10096-013-1910-0. Epub 2013 Jun 19. Eur J Clin Microbiol Infect Dis. 2013. PMID: 23780696
-
Update on hepatitis B and C virus diagnosis.World J Virol. 2015 Nov 12;4(4):323-42. doi: 10.5501/wjv.v4.i4.323. World J Virol. 2015. PMID: 26568915 Free PMC article. Review.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Medical