Pneumatic muscle actuator for resistive exercise in microgravity: test with a leg model
- PMID: 20131658
- DOI: 10.3357/asem.2486.2010
Pneumatic muscle actuator for resistive exercise in microgravity: test with a leg model
Abstract
Introduction: A proof-of-concept demonstration is described in which a DC servomotor (simulating the quadriceps of a human operator) rotated a pulley 90 degrees (simulating knee extension). A pneumatic muscle actuator (PMA) generated an opposing force (antagonist) to the rotating pulley. One application of such a device is for use in microgravity environments because the PMA is compact, simple, and of relatively small mass (283 g). In addition, the operator can set a computer-controlled force-level range in response to individual user changes in exercise conditioning over time.
Methods: A PMA was used in this study and interacted with a DC servomotor. For each trial, the PMA contracted in response to internal pressure. An input voltage profile activated the DC servomotor, resulting in the following three phases: an isokinetic counterclockwise pulley rotation of 90 degrees over 5 s (Phase I), the position was held for 5 s (Phase II), and an isokinetic clockwise rotation of 90 degrees over 5 s (Phase III). Root mean square error (RMSE) values were used to evaluate the pulley rotation.
Results: For Phase I, when the PMA pressures (in kPa) were 300, 450, and 575, the percent RMSE, respectively, were 5.24, 6.23, and 4.59. For Phase II, the percent RMSE were 2.81, 2.57, and 5.63, respectively. For Phase III, the percent RMSE were 5.69, 2.63, and 3.30, respectively.
Discussion: This study presents a demonstration of a PMA device that can enhance exercise by providing a wide range of resistive loads.
Similar articles
-
Pneumatic muscle actuator (PMA) task-specific resistance for potential use in microgravity exercise.Aviat Space Environ Med. 2012 Jul;83(7):696-701. doi: 10.3357/asem.3105.2012. Aviat Space Environ Med. 2012. PMID: 22779314
-
Characterisation of a pneumatic muscle test station with two dynamic plants in cascade.Comput Methods Biomech Biomed Engin. 2010 Feb;13(1):11-8. doi: 10.1080/10255842.2010.490087. Comput Methods Biomech Biomed Engin. 2010. PMID: 19488911
-
Haptic control of a pneumatic muscle actuator to provide resistance for simulated isokinetic exercise; part II: control development and testing.Comput Methods Biomech Biomed Engin. 2015;18(1):1-14. doi: 10.1080/10255842.2013.774380. Epub 2013 Mar 15. Comput Methods Biomech Biomed Engin. 2015. PMID: 23495753
-
Haptic control of a pneumatic muscle actuator to provide resistance for simulated isokinetic exercise: Part I--dynamic test station and human quadriceps dynamic simulator.Comput Methods Biomech Biomed Engin. 2014;17(12):1391-401. doi: 10.1080/10255842.2012.751982. Epub 2013 Jan 3. Comput Methods Biomech Biomed Engin. 2014. PMID: 23282135
-
Contributions to the understanding of gait control.Dan Med J. 2014 Apr;61(4):B4823. Dan Med J. 2014. PMID: 24814597 Review.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources