Gated treatment delivery verification with on-line megavoltage fluoroscopy
- PMID: 20133069
- DOI: 10.1016/j.ijrobp.2009.08.005
Gated treatment delivery verification with on-line megavoltage fluoroscopy
Abstract
Purpose: To develop and clinically demonstrate the use of on-line real-time megavoltage (MV) fluoroscopy for gated treatment delivery verification.
Methods and materials: Megavoltage fluoroscopy (MVF) image sequences were acquired using a flat panel equipped for MV cone-beam CT in synchrony with the respiratory signal obtained from the Anzai gating device. The MVF images can be obtained immediately before or during gated treatment delivery. A prototype software tool (named RTReg4D) was developed to register MVF images with phase-sequenced digitally reconstructed radiograph images generated from the treatment planning system based on four-dimensional CT. The image registration can be used to reposition the patient before or during treatment delivery. To demonstrate the reliability and clinical usefulness, the system was first tested using a thoracic phantom and then prospectively in actual patient treatments under an institutional review board-approved protocol.
Results: The quality of the MVF images for lung tumors is adequate for image registration with phase-sequenced digitally reconstructed radiographs. The MVF was found to be useful for monitoring inter- and intrafractional variations of tumor positions. With the planning target volume contour displayed on the MVF images, the system can verify whether the moving target stays within the planning target volume margin during gated delivery.
Conclusions: The use of MVF images was found to be clinically effective in detecting discrepancies in tumor location before and during respiration-gated treatment delivery. The tools and process developed can be useful for gated treatment delivery verification.
Similar articles
-
A technique of using gated-CT images to determine internal target volume (ITV) for fractionated stereotactic lung radiotherapy.Radiother Oncol. 2006 Feb;78(2):177-84. doi: 10.1016/j.radonc.2005.11.012. Epub 2006 Jan 10. Radiother Oncol. 2006. PMID: 16376444
-
Verification of respiratory-gated radiotherapy with new real-time tumour-tracking radiotherapy system using cine EPID images and a log file.Phys Med Biol. 2017 Feb 21;62(4):1585-1599. doi: 10.1088/1361-6560/aa587d. Epub 2017 Jan 10. Phys Med Biol. 2017. PMID: 28072584
-
Use of megavoltage cine-images for studying intra-thoracic motion during radiotherapy for locally advanced lung cancer.Radiother Oncol. 2011 May;99(2):155-60. doi: 10.1016/j.radonc.2011.05.037. Epub 2011 May 27. Radiother Oncol. 2011. PMID: 21621867
-
Integrating respiratory gating into a megavoltage cone-beam CT system.Med Phys. 2006 Jul;33(7):2354-61. doi: 10.1118/1.2207136. Med Phys. 2006. PMID: 16898437
-
The adaptation of megavoltage cone beam CT for use in standard radiotherapy treatment planning.Phys Med Biol. 2009 Apr 7;54(7):2067-77. doi: 10.1088/0031-9155/54/7/014. Epub 2009 Mar 13. Phys Med Biol. 2009. PMID: 19287087
Cited by
-
Intrafraction verification of gated RapidArc by using beam-level kilovoltage X-ray images.Int J Radiat Oncol Biol Phys. 2012 Aug 1;83(5):e709-15. doi: 10.1016/j.ijrobp.2012.03.006. Epub 2012 May 2. Int J Radiat Oncol Biol Phys. 2012. PMID: 22554582 Free PMC article.
-
Use of dMLC for implementation of dynamic respiratory-gated radiation therapy.Med Phys. 2013 Oct;40(10):101708. doi: 10.1118/1.4820534. Med Phys. 2013. PMID: 24089898 Free PMC article.
-
Clinical implementation of intrafraction cone beam computed tomography imaging during lung tumor stereotactic ablative radiation therapy.Int J Radiat Oncol Biol Phys. 2013 Dec 1;87(5):917-23. doi: 10.1016/j.ijrobp.2013.08.015. Epub 2013 Oct 8. Int J Radiat Oncol Biol Phys. 2013. PMID: 24113060 Free PMC article.
-
Motion management strategies and technical issues associated with stereotactic body radiotherapy of thoracic and upper abdominal tumors: A review from NRG oncology.Med Phys. 2017 Jun;44(6):2595-2612. doi: 10.1002/mp.12227. Epub 2017 Apr 20. Med Phys. 2017. PMID: 28317123 Free PMC article.
-
Evaluation of the geometric accuracy of surrogate-based gated VMAT using intrafraction kilovoltage x-ray images.Med Phys. 2012 May;39(5):2686-93. doi: 10.1118/1.4704729. Med Phys. 2012. PMID: 22559639 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources