Correlation between laser fluorescence readings and volume of tooth preparation in incipient occlusal caries in vitro
- PMID: 20136944
- DOI: 10.1111/j.1708-8240.2009.00309.x
Correlation between laser fluorescence readings and volume of tooth preparation in incipient occlusal caries in vitro
Abstract
Objective: This study evaluated the correlation between laser fluorescence readings and the extent of incipient occlusal caries as measured by the volume of tooth preparation in vitro.
Materials and methods: One hundred and three permanent molars and premolars containing incipient occlusal pit-and-fissure caries and sound occlusal surfaces (1/4 of the sample, control) were selected. DIAGNOdent (KaVo Dental Corporation, Lake Zurich, IL, USA) readings were obtained according to manufacturer instructions. Caries was removed with 1/4 round burs in high speed. The volume of tooth preparation was measured using a surrogate measure based on the amount of composite needed to fill the preparations. Sensitivity and specificity using different cutoff values were calculated for lesions/preparations extending into dentin. The results were analyzed statistically.
Results: The Pearson correlation for preparation volume and DIAGNOdent reading measurements was low (r = 0.285). Sensitivity and specificity of DIAGNOdent for detection of dentinal lesions were 0.83 and 0.60, and 0.66 and 0.73 for the cutoff values of 20 and 30, respectively.
Conclusions: Within the limitations of this study, laser fluorescence measured with DIAGNOdent does not correlate well with extent of carious tooth structure in incipient occlusal caries.
Clinical significance: Clinicians should not rely only on DIAGNOdent readings to determine the extension of incipient occlusal caries.
Comment in
-
Commentary. Correlation between laser fluorescence readings and volume of tooth preparation in incipient occlusal caries in vitro.J Esthet Restor Dent. 2010 Feb;22(1):40-1. doi: 10.1111/j.1708-8240.2009.00310.x. J Esthet Restor Dent. 2010. PMID: 20136945 No abstract available.
Similar articles
-
Commentary. Correlation between laser fluorescence readings and volume of tooth preparation in incipient occlusal caries in vitro.J Esthet Restor Dent. 2010 Feb;22(1):40-1. doi: 10.1111/j.1708-8240.2009.00310.x. J Esthet Restor Dent. 2010. PMID: 20136945 No abstract available.
-
Comparison of different DIAGNOdent cut-off limits for in vivo detection of occlusal caries.Oper Dent. 2003 Nov-Dec;28(6):672-80. Oper Dent. 2003. PMID: 14653279
-
Comparison of two laser fluorescence devices for the detection of occlusal caries in vivo.Eur J Oral Sci. 2007 Aug;115(4):252-6. doi: 10.1111/j.1600-0722.2007.00456.x. Eur J Oral Sci. 2007. PMID: 17697163
-
Caries process on occlusal surfaces: evolving evidence and understanding.Caries Res. 2014;48(4):339-46. doi: 10.1159/000356307. Caries Res. 2014. PMID: 24577073 Review.
-
Clinical use of lasers in caries diagnosis and therapy.Ann R Australas Coll Dent Surg. 2008 Jun;19:62-5. Ann R Australas Coll Dent Surg. 2008. PMID: 22073455 Review.
Cited by
-
Caries diagnosis using light fluorescence devices: VistaProof and DIAGNOdent.Odontology. 2014 Jul;102(2):330-5. doi: 10.1007/s10266-013-0105-6. Epub 2013 Mar 7. Odontology. 2014. PMID: 23467922
-
Fluorescence devices for the detection of dental caries.Cochrane Database Syst Rev. 2020 Dec 8;12(12):CD013811. doi: 10.1002/14651858.CD013811. Cochrane Database Syst Rev. 2020. PMID: 33319353 Free PMC article.
-
Correlation with Caries Lesion Depth of The Canary System, DIAGNOdent and ICDAS II.Open Dent J. 2017 Dec 29;11:679-689. doi: 10.2174/1874210601711010679. eCollection 2017. Open Dent J. 2017. PMID: 29387284 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical