Lumbar axial loading device alters lumbar sagittal alignment differently from upright standing position: a computed tomography study
- PMID: 20139804
- DOI: 10.1097/BRS.0b013e3181bb8188
Lumbar axial loading device alters lumbar sagittal alignment differently from upright standing position: a computed tomography study
Abstract
Study design: A study was performed using an axial loading device in healthy young subjects.
Objective: To determine whether sagittal alignment during axial loading using a compression device can accurately simulate the standing posture.
Summary of background data: Axial compression devices are widely used for simulation of standing position during magnetic resonance imaging (MRI) or computed tomography (CT) scans. However, images taken during axial loading have not been compared with those obtained in a standing posture.
Methods: The study population comprised 14 asymptomatic healthy volunteers (7 men and 7 women: age 21-32, mean 27 years). Lumbar lateral radiograph films obtained in the standing posture (standing condition), lumbar CT images with axial loading using a DynaWell compression device (axial loading condition), and CT images without loading (control) were compared. Changes in spinal length, lumbar disc height, segmental lordotic angle, and total lumbar lordotic angle were compared among the conditions.
Results: Spinal length was significantly decreased in both the axial loading and standing conditions compared with controls. The magnitude of the changes was greater in the standing condition than in the axial loading condition. Segmental lordotic angle at L2/3 and L3/4 was significantly increased in both axial loading and standing conditions. However, disc lordotic angle at L5/S was significantly decreased in the axial loading condition, while the standing condition showed no significant change. Consequently, the pelvic angle showed a significant decrease in the axial loading condition.
Conclusion: The compression device simulates the lumbar segmental alignment change from supine to standing posture in L1/2, L2/3, L3/4, and L4/5. However, in L5/S, axial loading using the DynaWell altered lumbar segmental alignment with a kyphotic change, while no significant difference was observed in this level between standing and supine positions. Awareness of these phenomena are essential for accurate interpretation of imaging results.
Similar articles
-
Test-retest repeatability of lumbar sagittal alignment and disc height measurements with or without axial loading: a computed tomography study.J Spinal Disord Tech. 2011 Apr;24(2):93-8. doi: 10.1097/BSD.0b013e3181dd611f. J Spinal Disord Tech. 2011. PMID: 21430497
-
In vivo measurement of lumbar foramen during axial loading using a compression device and computed tomography.J Spinal Disord Tech. 2013 Jul;26(5):E177-82. doi: 10.1097/BSD.0b013e318286f635. J Spinal Disord Tech. 2013. PMID: 23381186
-
Lumbar spine disc height and curvature responses to an axial load generated by a compression device compatible with magnetic resonance imaging.Spine (Phila Pa 1976). 2001 Dec 1;26(23):2596-600. doi: 10.1097/00007632-200112010-00014. Spine (Phila Pa 1976). 2001. PMID: 11725241
-
Assessment of lower limb alignment: supine weight-bearing CT scanograms compared with a standing full-length radiograph.Skeletal Radiol. 2024 Aug;53(8):1465-1471. doi: 10.1007/s00256-024-04637-z. Epub 2024 Mar 5. Skeletal Radiol. 2024. PMID: 38443696 Review.
-
Weight-bearing MRI of the Lumbar Spine: Technical Aspects.Semin Musculoskelet Radiol. 2019 Dec;23(6):609-620. doi: 10.1055/s-0039-1697936. Epub 2019 Nov 19. Semin Musculoskelet Radiol. 2019. PMID: 31745951 Review.
Cited by
-
Dynamic MR in patients affected by neurogenical claudication: technique and results from a single-center experience.Neuroradiology. 2016 Aug;58(8):765-70. doi: 10.1007/s00234-016-1697-7. Epub 2016 May 21. Neuroradiology. 2016. PMID: 27209418
-
Axial loading lumbar magnetic resonance imaging with a new device in asymptomatic individuals.Quant Imaging Med Surg. 2023 Jan 1;13(1):58-65. doi: 10.21037/qims-22-283. Epub 2022 Sep 22. Quant Imaging Med Surg. 2023. PMID: 36620162 Free PMC article.
-
Evaluation of Subtalar Joint's Compensatory Function in Varus Ankle Osteoarthritis Using Globally Optimal Iterative Closest Points (Go-ICP).Foot Ankle Orthop. 2022 Jun 27;7(2):24730114221103584. doi: 10.1177/24730114221103584. eCollection 2022 Apr. Foot Ankle Orthop. 2022. PMID: 35782686 Free PMC article.
-
High-resolution, three-dimensional magnetic resonance imaging axial load dynamic study improves diagnostics of the lumbar spine in clinical practice.World J Orthop. 2022 Jan 18;13(1):87-101. doi: 10.5312/wjo.v13.i1.87. eCollection 2022 Jan 18. World J Orthop. 2022. PMID: 35096539 Free PMC article.
-
Disc deformation as a potential biomarker of nonspecific low back pain.Sci Rep. 2025 Aug 5;15(1):28480. doi: 10.1038/s41598-025-14366-5. Sci Rep. 2025. PMID: 40759706 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical
Research Materials