Finite element analysis of the effects of pedicle screw fixation nut loosening on lumbar interbody fusion based on the elasto-plateau plasticity of bone characteristics
- PMID: 20139810
- DOI: 10.1097/BRS.0b013e3181b6258a
Finite element analysis of the effects of pedicle screw fixation nut loosening on lumbar interbody fusion based on the elasto-plateau plasticity of bone characteristics
Abstract
Study design: Cylindrical cages and pedicle screw fixation (PSF) have been used for the setting of a lumbar interbody fusion.
Objective: To investigate the effects of PSF-nut loosening on lumbar interbody fusion based on the elasto-plateau plasticity of bone characteristics.
Summary of background data: The stress-strain curves of bone materials beneath the cages show linear elastic behavior before yielding and then seem to be plateau plastic deformation with further compression strain until opposing cell walls contact.
Methods: Cancellous bone was assumed to be linearly elastic followed by nonlinearly plastic such as crushable foam behavior. The deformation and failure processes caused by loosening of PSF-nuts, in particular, are addressed by means of finite element analysis. Two finite element models, including anterior (or posterior) lumbar interbody fusion for the human lumbar L4-L5 segments, were constructed. In the respective models, 2 types of PSF were considered such as a solid PSF without relative movement between rod and screw, and a tightened PSF with relative axial movement of the rod between them due to the slight loosening of 2 nuts of L4.
Results: The slight loosening of nuts showed much greater levels of bone strain with relative motion at the bone-implant interface compared with the PSF with firmly tightened nuts. The loosening 2 of the 4 nuts imposed an excessive burden on the remaining screws.
Conclusion: By the consideration of the elasto-plateau plastic behavior of cancellous bone and the relative motion at the contact interface between screw and rod, much more exact behaviors of bone failure were predicted. Slight loosening of the nut would cause a possible risk of segmental instability. With difficulty in either tightening the nut firmly during actual operation or keeping the nut-locking mechanism functioning perfectly in the postoperative state, an enhanced design methodology with more effective locking mechanism is required. Slight PSF-nut loosening may increase risks of failure processes such as rod migration, rod breakage, screw breakage, screw loosening or cage loosening. Different mechanisms however, can be considered; rod breakage and screw breakages are related to the higher stress level of the rod and screw while rod migration is associated with slight nut-loosening. Especially, in the case of a nonsymmetric nut-loosening the risk would be increased. Thus, further study is necessary for improving the current design of PSFs.
Similar articles
-
Finite element analysis of anterior lumbar interbody fusion: threaded cylindrical cage and pedicle screw fixation.Spine (Phila Pa 1976). 2007 Nov 1;32(23):2558-68. doi: 10.1097/BRS.0b013e318158cdd8. Spine (Phila Pa 1976). 2007. PMID: 17978654
-
Bilateral pedicle screw fixation provides superior biomechanical stability in transforaminal lumbar interbody fusion: a finite element study.Spine J. 2015 Aug 1;15(8):1812-22. doi: 10.1016/j.spinee.2014.06.015. Epub 2014 Jun 28. Spine J. 2015. PMID: 24983669
-
Biomechanical assessment of anterior lumbar interbody fusion with an anterior lumbosacral fixation screw-plate: comparison to stand-alone anterior lumbar interbody fusion and anterior lumbar interbody fusion with pedicle screws in an unstable human cadaver model.Spine (Phila Pa 1976). 2006 Apr 1;31(7):762-8. doi: 10.1097/01.brs.0000206360.83728.d2. Spine (Phila Pa 1976). 2006. PMID: 16582849
-
Are stand-alone cages sufficient for anterior lumbar interbody fusion?Orthop Surg. 2012 Feb;4(1):11-4. doi: 10.1111/j.1757-7861.2011.00164.x. Orthop Surg. 2012. PMID: 22290813 Free PMC article. Review.
-
Prediction of Pedicle Screw Loosening Using an MRI-Based Vertebral Bone Quality Score in Patients with Lumbar Degenerative Disease.World Neurosurg. 2023 Mar;171:e760-e767. doi: 10.1016/j.wneu.2022.12.098. Epub 2022 Dec 27. World Neurosurg. 2023. PMID: 36584889 Review.
Cited by
-
Comparison of short-segment monoaxial and polyaxial pedicle screw fixation combined with intermediate screws in traumatic thoracolumbar fractures: a finite element study and clinical radiographic review.Clinics (Sao Paulo). 2017 Oct;72(10):609-617. doi: 10.6061/clinics/2017(10)04. Clinics (Sao Paulo). 2017. PMID: 29160423 Free PMC article.
-
Treatment of unstable thoracolumbar fractures through short segment pedicle screw fixation techniques using pedicle fixation at the level of the fracture: a finite element analysis.PLoS One. 2014 Jun 10;9(6):e99156. doi: 10.1371/journal.pone.0099156. eCollection 2014. PLoS One. 2014. PMID: 24914815 Free PMC article.
-
Hybrid pedicle screw and modified cortical bone trajectory technique in transforaminal lumbar interbody fusion at L4-L5 segment: finite element analysis.BMC Musculoskelet Disord. 2023 Apr 13;24(1):288. doi: 10.1186/s12891-023-06385-y. BMC Musculoskelet Disord. 2023. PMID: 37055739 Free PMC article.
-
Effect of the In Situ Screw Implantation Region and Angle on the Stability of Lateral Lumbar Interbody Fusion: A Finite Element Study.Orthop Surg. 2022 Jul;14(7):1506-1517. doi: 10.1111/os.13312. Epub 2022 Jun 3. Orthop Surg. 2022. PMID: 35656700 Free PMC article.
-
Finite Element Analysis of a Novel Anterior Locking Plate for Thoracolumbar Burst Fracture.Biomed Res Int. 2021 Oct 11;2021:2949419. doi: 10.1155/2021/2949419. eCollection 2021. Biomed Res Int. 2021. PMID: 34671672 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Research Materials