Image fusion in dual energy computed tomography for detection of hypervascular liver hepatocellular carcinoma: phantom and preliminary studies
- PMID: 20142749
- DOI: 10.1097/RLI.0b013e3181d32119
Image fusion in dual energy computed tomography for detection of hypervascular liver hepatocellular carcinoma: phantom and preliminary studies
Abstract
Objective: This study was designed to determine the optimal blending method and parameters to fuse computed tomography (CT) data sets with different energy levels in dual-energy CT (DECT) for the detection of hypervascular liver lesions.
Materials and methods: A liver agar phantom containing 8 conical tubes with various concentrations of contrast material, was scanned using a Somatom Definition Dual Source CT (DSCT; Siemens, Forchheim, Germany) scanner in the dual energy mode at different current settings. CT data sets obtained at voltage potentials of 80 kVp and 140 kVp were fused using the linear blending method and nonlinear method with different weighting factors (0.1, 0.3, 0.5, 0.7, and 0.9) and different parameters sets (A--lambda: 20, omega: 430; B--lambda: 20, omega: 70; C--lambda: 250, omega: 430; D--lambda: 250, omega: 70). In 20 patients with hepatocellular carcinomas, multiphasic liver CT scans including arterial, portal, and equilibrium phases were performed. DECT was used only during the arterial phase but a voltage potential of 120 kVp was used for both the portal and equilibrium phases. For quantitative analyses of the phantom and patient study, the contrast-to-noise ratio (CNR) of the lesion to liver on arterial phase images, was measured. For qualitative analysis of the CT images of the 20 study patients, 5 radiologists, each with a different level of clinical experience, independently assessed the 5 types of image sets regarding lesion conspicuity and overall image quality. This study followed the guidelines of our hospital's institutional review board, and patient informed written consent was not required. Statistical comparisons were made using repeated measures ANOVA with Bonferroni correction for multiple comparisons.
Results: For the phantom and patient studies, 2 linear images with weighting factors 0.5 and 0.7 and 2 nonlinear images with a wide width, showed a higher CNR of hyperattenuated lesions than a standard 0.3 weighting factor linear blended image (P < 0.05). For the patient study, a weighting factor 0.5 and a 0.7 linear mixing image had a 9.2% and an 11.8% increase in CNR, respectively, more than a 0.3 linear blended image (P < 0.05). Moreover, a nonlinear mixing image with parameters A and C had a 14.0% increase in CNR over that of a 0.3 linear blended image (P < 0.05). In a qualitative study performed by the 5 reviewers, the nonlinear blended image set with a low level and wide width, was estimated as the most preferred image set (55%-100%), whereas a weighting factor of a 0.3 linear blended image was determined as the least preferred image (65%-100%).
Conclusions: Linear blended images with a higher weighting factor than 0.5 and nonlinear blended images with a wide width are able to provide improved lesion-to-liver CNR over the standard linear blending method (with a weighting factor of 0.3). Furthermore, the linear blending method with a low level and a high width provided the most preferred image set for hypervascular hepatocellular carcinoma detection by our radiologists.
Similar articles
-
Image quality of mean temporal arterial and mean temporal portal venous phase images calculated from low dose dynamic volume perfusion CT datasets in patients with hepatocellular carcinoma and pancreatic cancer.Eur J Radiol. 2016 Nov;85(11):2104-2110. doi: 10.1016/j.ejrad.2016.09.024. Epub 2016 Sep 25. Eur J Radiol. 2016. PMID: 27776665
-
80-kVp CT using Iterative Reconstruction in Image Space algorithm for the detection of hypervascular hepatocellular carcinoma: phantom and initial clinical experience.Korean J Radiol. 2012 Mar-Apr;13(2):152-64. doi: 10.3348/kjr.2012.13.2.152. Epub 2012 Mar 7. Korean J Radiol. 2012. PMID: 22438682 Free PMC article.
-
Low tube voltage intermediate tube current liver MDCT: sinogram-affirmed iterative reconstruction algorithm for detection of hypervascular hepatocellular carcinoma.AJR Am J Roentgenol. 2013 Jul;201(1):23-32. doi: 10.2214/AJR.12.10000. AJR Am J Roentgenol. 2013. PMID: 23789655
-
Hepatocellular nodules in liver cirrhosis: state of the art CT evaluation (perfusion CT/volume helical shuttle scan/dual-energy CT, etc.).Abdom Imaging. 2011 Jun;36(3):273-81. doi: 10.1007/s00261-011-9684-2. Abdom Imaging. 2011. PMID: 21267563 Review.
-
CT findings of primary clear cell carcinoma of liver: with analysis of 19 cases and review of the literature.Abdom Imaging. 2014 Aug;39(4):736-43. doi: 10.1007/s00261-014-0104-2. Abdom Imaging. 2014. PMID: 24549879 Review.
Cited by
-
Comparison of dual-energy computed tomography (DECT) polychromatic and monochromatic images with and without iterative metal artifact reduction algorithm in patients with dental implants.Quant Imaging Med Surg. 2024 Jul 1;14(7):4688-4702. doi: 10.21037/qims-24-19. Epub 2024 Jun 27. Quant Imaging Med Surg. 2024. PMID: 39022239 Free PMC article.
-
Image quality and radiation dose of dual-energy CT of the head and neck compared with a standard 120-kVp acquisition.AJNR Am J Neuroradiol. 2011 Dec;32(11):1994-9. doi: 10.3174/ajnr.A2654. Epub 2011 Sep 8. AJNR Am J Neuroradiol. 2011. PMID: 21903909 Free PMC article.
-
Comparative study of hepatic venography using non-linear-blending images, monochromatic images and low-voltage images of dual-energy CT.Br J Radiol. 2014 Oct;87(1042):20140220. doi: 10.1259/bjr.20140220. Epub 2014 Jul 23. Br J Radiol. 2014. PMID: 25051976 Free PMC article.
-
Optofluidic phantom mimicking optical properties of porcine livers.Biomed Opt Express. 2011 Jul 1;2(7):1877-92. doi: 10.1364/BOE.2.001877. Epub 2011 Jun 9. Biomed Opt Express. 2011. PMID: 21750766 Free PMC article.
-
Effects of High-concentration contrast material and low-voltage CT on contrast for multiphasic CT of the upper abdomen: comparison using the simulation with virtual monochromatic imaging obtained by fast-switch kVp dual-energy CT.Springerplus. 2014 May 8;3:234. doi: 10.1186/2193-1801-3-234. eCollection 2014. Springerplus. 2014. PMID: 24891998 Free PMC article.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical