Usability, participant acceptance, and safety of a prefilled insulin injection device in a 3-month observational survey in everyday clinical practice in Australia
- PMID: 20144398
- PMCID: PMC2787044
- DOI: 10.1177/193229680900300624
Usability, participant acceptance, and safety of a prefilled insulin injection device in a 3-month observational survey in everyday clinical practice in Australia
Abstract
Background: SoloSTAR (SOL; sanofi-aventis, Paris, France) is a prefilled insulin pen device for the injection of insulin glargine and insulin glulisine. This is the first Australian survey to determine its usability, participant acceptance, and safety in clinical practice.
Methods: A 3-month, nonrandomized, noncomparative, observational survey in Australia was conducted in individuals with diabetes. Participants were given SOL pens containing glargine, the instruction leaflet, and a toll-free helpline number. Training was offered to all participants. Safety data, including product technical complaints (PTCs), were gathered from ongoing feedback given by the participant or health care professional (HCP) and by independent interviews conducted 6-10 weeks after study start.
Results: Some 2674 people consented to take part across 93 sites (150 HCPs), and 2029 participated in interviews. Of these, 52.6% had type 1 diabetes, 16.3% had manual dexterity problems, and 15.5% had poor eyesight not corrected by glasses. At the time of interview, 96.8% of participants were still using SOL. None of the eight PTCs reported were due to technical defects; most were related to handling errors. Some 62 participants reported 77 adverse events; none were related to a PTC. The vast majority of participants (95.4%) were "very satisfied" or "satisfied" with using SOL, and 89.7% of the participants had no questions or concerns using SOL on a daily basis. Similar positive findings were reported by participants with manual or dexterity impairments.
Conclusions: In this survey of everyday clinical practice, SOL had a good safety profile and was very well accepted by participants.
Figures
Comment on
-
An analysis of patient acceptance and safety of a prefilled insulin injection device.J Diabetes Sci Technol. 2009 Nov 1;3(6):1439-41. doi: 10.1177/193229680900300625. J Diabetes Sci Technol. 2009. PMID: 20144399 Free PMC article.
Similar articles
-
An analysis of patient acceptance and safety of a prefilled insulin injection device.J Diabetes Sci Technol. 2009 Nov 1;3(6):1439-41. doi: 10.1177/193229680900300625. J Diabetes Sci Technol. 2009. PMID: 20144399 Free PMC article.
-
Usability of a pre-filled insulin injection device in a 3-month observational survey of everyday clinical practice in Australia.Curr Med Res Opin. 2008 Oct;24(10):2741-9. doi: 10.1185/03007990802367579. Epub 2008 Aug 19. Curr Med Res Opin. 2008. PMID: 18715525
-
[Evaluation of a new disposable insulin pen and injection habits of diabetes patients in everyday clinical practice].MMW Fortschr Med. 2010 Jan 14;151 Suppl 4:179-87. MMW Fortschr Med. 2010. PMID: 21595146 Clinical Trial. German.
-
Dose accuracy and injection force dynamics of a novel disposable insulin pen.Expert Opin Drug Deliv. 2007 Mar;4(2):165-74. doi: 10.1517/17425247.4.2.165. Expert Opin Drug Deliv. 2007. PMID: 17335413 Review.
-
Insulin glargine and glulisine SoloSTAR pens for the treatment of diabetes.Expert Rev Med Devices. 2008 Mar;5(2):113-23. doi: 10.1586/17434440.5.2.113. Expert Rev Med Devices. 2008. PMID: 18331174 Review.
Cited by
-
Formative usability evaluation of a fixed-dose pen-injector platform device.Med Devices (Auckl). 2018 Apr 4;11:105-112. doi: 10.2147/MDER.S159733. eCollection 2018. Med Devices (Auckl). 2018. PMID: 29670411 Free PMC article.
-
Usability of devices for self-injection: results of a formative study on a new disposable pen injector.Med Devices (Auckl). 2014 Jun 12;7:195-203. doi: 10.2147/MDER.S63918. eCollection 2014. Med Devices (Auckl). 2014. PMID: 24966698 Free PMC article.
-
Evaluation of Patient Satisfaction After Using Prefilled Insulin Pen in Real Life in Lithuania.J Diabetes Sci Technol. 2015 Jul;9(4):939-40. doi: 10.1177/1932296815577426. Epub 2015 Mar 20. J Diabetes Sci Technol. 2015. PMID: 25795640 Free PMC article. No abstract available.
-
An analysis of patient acceptance and safety of a prefilled insulin injection device.J Diabetes Sci Technol. 2009 Nov 1;3(6):1439-41. doi: 10.1177/193229680900300625. J Diabetes Sci Technol. 2009. PMID: 20144399 Free PMC article.
-
A Pan-European and Canadian prospective survey to evaluate patient satisfaction with the SoloSTAR insulin injection device in type 1 and type 2 diabetes.J Diabetes Sci Technol. 2011 Sep 1;5(5):1224-34. doi: 10.1177/193229681100500531. J Diabetes Sci Technol. 2011. PMID: 22027323 Free PMC article.
References
-
- Bohannon NJ. Insulin delivery using pen devices. Simple-to-use tools may help young and old alike. Postgrad Med. 1999;106(5):57–58. 61–4, 68. - PubMed
-
- Graff MR, McClanahan MA. Assessment by patients with diabetes mellitus of two insulin pen delivery systems versus a vial and syringe. Clin Ther. 1998;20(3):486–496. - PubMed
-
- Klonoff DC. The pen is mightier than the needle (and syringe) Diabetes Technol Ther. 2001;3(4):631–633. - PubMed
-
- Summers KH, Szeinbach SL, Lenox SM. Preference for insulin delivery systems among current insulin users and nonusers. Clin Ther. 2004;26(9):1498–1505. - PubMed
-
- Clarke A, Dain MP. Dose accuracy of a reusable insulin pen using a cartridge system with an integrated plunger mechanism. Expert Opin Drug Deliv. 2006;3(5):677–683. - PubMed
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Medical
Miscellaneous