Accuracy of the SEVEN continuous glucose monitoring system: comparison with frequently sampled venous glucose measurements
- PMID: 20144429
- PMCID: PMC2769895
- DOI: 10.1177/193229680900300519
Accuracy of the SEVEN continuous glucose monitoring system: comparison with frequently sampled venous glucose measurements
Abstract
Background: The purpose of this study was to compare the accuracy of measurements obtained from the DexCom SEVEN system with Yellow Springs Instrument (YSI) laboratory measurements of venous blood glucose.
Methods: Seventy-two subjects with insulin-requiring diabetes, aged 18-71, were enrolled in a multicenter, prospective single-arm study. All participants wore the SEVEN continuous glucose monitoring (CGM) system for one, 7-day wear period. Calibration with capillary finger stick measurements was performed 2 hours after sensor insertion and once every 12 hours thereafter. A subset of subjects (28) wore two systems simultaneously to assess precision. All subjects participated in one, 10-hour in-clinic session on day 1, 4, or 7 of the study to compare CGM measurements against a laboratory method (YSI analyzer) using venous measurements taken once every 20 minutes. Carbohydrate consumption and insulin dosing were adjusted in order to obtain a broad range of glucose values.
Results: Comparison of CGM measurements with the laboratory reference method (n = 2318) gave mean and median absolute relative differences (ARDs) of 16.7 and 13.2%, respectively. The percentage was 70.4% in the clinically accurate Clarke error grid A zone and 27.5% in the benign error B zone. Performance of the SEVEN system was consistent over time with mean and median ARD lowest on day 7 as compared to YSI (13.3 and 10.2%, respectively). Average sensor time lag was 5 minutes.
Conclusions: Measurements of the DexCom SEVEN system were found to be consistent and accurate compared with venous measurements made using a laboratory reference method over 7 days of wear.
2009 Diabetes Technology Society.
Figures
Comment on
-
An analysis of the SEVEN system: have we reached the summit of needle-type sensor accuracy?J Diabetes Sci Technol. 2009 Sep 1;3(5):1155-7. doi: 10.1177/193229680900300520. J Diabetes Sci Technol. 2009. PMID: 20144430 Free PMC article.
References
-
- Skyler JS. The economic burden of diabetes and the benefits of improved glycemic control: the potential role of a continuous glucose monitoring system. Diabetes Technol Ther. 2000;2(Suppl. 1):S7–S12. - PubMed
-
- Klonoff DC. Continuous glucose monitoring: roadmap for 21st century diabetes therapy. Diabetes Care. 2005;28(5):1231–1239. - PubMed
-
- Buckingham B, Block J, Wilson DM. Continuous glucose monitoring. Curr Opin Endocrinol Diabetes. 2005;12:273–279. - PubMed
-
- Bailey T, Zisser H, Garg S. Reduction in reduction in hemoglobin A1c with real-time continuous glucose monitoring: results from a 12-week observational study. Diabetes Technol Ther. 2007;9(3):203–209. - PubMed
-
- Garg S, Jovanovic L. Relationship of fasting and hourly blood glucose levels to HbA1c values: safety, accuracy, and improvements in glucose profiles obtained using a 7-day continuous glucose sensor. Diabetes Care. 2005;29(12):2644–2649. - PubMed
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Medical