Evaluating imaging systems: practical applications
- PMID: 20147386
- DOI: 10.1093/rpd/ncq007
Evaluating imaging systems: practical applications
Abstract
There are many ways in which imaging systems can be evaluated. The aim of the present paper is to provide an overview of a number of selected approaches to evaluating imaging systems, often encountered by the medical physicist, and discuss their validity and reliability. Specifically, it will cover (i) characterisation of an imaging system in terms of its detective quantum efficiency using linear-systems analysis; (ii) attempts to calculate relevant measures directly in images using the Rose model and the pixel signal-to-noise ratio; (iii) task-based methods incorporating human observers such as receiver-operating characteristics and (iv) visual grading-based methods using experienced radiologists as observers.
Similar articles
-
256-channel magnetic imaging system.Rev Sci Instrum. 2008 Jan;79(1):013709. doi: 10.1063/1.2829883. Rev Sci Instrum. 2008. PMID: 18248041
-
Mumford and Shah functional: VLSI analysis and implementation.IEEE Trans Pattern Anal Mach Intell. 2006 Mar;28(3):487-94. doi: 10.1109/TPAMI.2006.59. IEEE Trans Pattern Anal Mach Intell. 2006. PMID: 16526435
-
Multistatic synthetic aperture radar image formation.IEEE Trans Image Process. 2010 May;19(5):1290-306. doi: 10.1109/TIP.2009.2039662. Epub 2009 Dec 31. IEEE Trans Image Process. 2010. PMID: 20051343
-
Systems for research and development in medical ultrasound imaging.Med Device Technol. 2004 Jun;15(5):35-7. Med Device Technol. 2004. PMID: 15285488 Review.
-
Autocalibrated coil sensitivity estimation for parallel imaging.NMR Biomed. 2006 May;19(3):316-24. doi: 10.1002/nbm.1048. NMR Biomed. 2006. PMID: 16705632 Review.
Cited by
-
Dependency of image quality on acquisition protocol and image processing in chest tomosynthesis-a visual grading study based on clinical data.Br J Radiol. 2018 Jul;91(1087):20170683. doi: 10.1259/bjr.20170683. Epub 2018 Apr 9. Br J Radiol. 2018. PMID: 29565673 Free PMC article.
-
Evaluation of three presets for four-dimensional cone beam CT in lung radiotherapy verification by visual grading analysis.Br J Radiol. 2016 Jul;89(1063):20150933. doi: 10.1259/bjr.20150933. Epub 2016 Apr 25. Br J Radiol. 2016. PMID: 27109735 Free PMC article.
-
Image quality in ultra-low-dose chest CT versus chest x-rays guiding paediatric cystic fibrosis care.Eur Radiol. 2025 Jul 25. doi: 10.1007/s00330-025-11835-3. Online ahead of print. Eur Radiol. 2025. PMID: 40711551
-
Impact of deep learning model uncertainty on manual corrections to MRI-based auto-segmentation in prostate cancer radiotherapy.J Appl Clin Med Phys. 2025 Sep;26(9):e70221. doi: 10.1002/acm2.70221. J Appl Clin Med Phys. 2025. PMID: 40849835 Free PMC article.
-
Visual grading analysis of digital neonatal chest phantom X-ray images: Impact of detector type, dose and image processing on image quality.Eur Radiol. 2018 Jul;28(7):2951-2959. doi: 10.1007/s00330-017-5301-2. Epub 2018 Feb 19. Eur Radiol. 2018. PMID: 29460076
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources