Bilevel positive airway pressure ventilation: factors influencing carbon dioxide rebreathing
- PMID: 20148321
- DOI: 10.1007/s00134-010-1774-z
Bilevel positive airway pressure ventilation: factors influencing carbon dioxide rebreathing
Abstract
Purpose: Use of bilevel positive airway pressure (BLPAP) ventilators for noninvasive ventilation (NIV) is an established treatment for both acute and chronic ventilatory failure. Although BLPAP ventilator circuits are simpler than those of conventional ventilators, one drawback to their use is that they allow variable amounts of rebreathing to occur. The aim of this work is to measure the amount of CO(2) reinsufflated in relation to the BLPAP ventilator circuit in patients, and to determine predictive factors for rebreathing.
Methods: Eighteen adult patients were ventilated on pressure support, either by intubation or with mask ventilation, during a weaning period. The mean inspiratory fraction of CO(2) (tidal FiCO(2)) reinsufflated from the circuit between the intentional leak and the ventilator was measured for each breath. The influence of end-tidal CO(2) concentration (ETCO(2)), respiratory rate (RR), percentage of inspiratory time (T (i)/T (TOT)), application of expiratory positive airway pressure (EPAP), and inspiratory tidal volume on magnitude of tidal FiCO(2), as well as the influence of intubation versus NIV, were studied by univariate comparisons and logistic regression analysis.
Results: In a total of 11,107 cycles, tidal FiCO(2) was 0.072 +/- 0.173%. Of fractions measured, 8,976 (81%) were under 0.10% and 2,131 (19%) were over 0.10%, with mean values of 0.026 +/- 0.027% and 0.239 +/- 0.326%, respectively. ETCO(2), EPAP, NIV versus intubation, and RR had significant predictive value for tidal FiCO(2) >0.10%.
Conclusions: BLPAP ventilators present a specific rebreathing risk to patients. However, that risk remains modest, even in intubated patients, provided that EPAP is applied.
Similar articles
-
Evaluation of carbon dioxide rebreathing during pressure support ventilation with airway management system (BiPAP) devices.Chest. 1995 Sep;108(3):772-8. doi: 10.1378/chest.108.3.772. Chest. 1995. PMID: 7656632 Clinical Trial.
-
Position of exhalation port and mask design affect CO2 rebreathing during noninvasive positive pressure ventilation.Crit Care Med. 2003 Aug;31(8):2178-82. doi: 10.1097/01.CCM.0000081309.71887.E9. Crit Care Med. 2003. PMID: 12973177
-
A comparison of controlled ventilation with a noninvasive ventilator versus traditional mask ventilation.J Clin Monit Comput. 2020 Aug;34(4):771-777. doi: 10.1007/s10877-019-00365-1. Epub 2019 Jul 23. J Clin Monit Comput. 2020. PMID: 31338661
-
Equipment needs for noninvasive mechanical ventilation.Eur Respir J. 2002 Oct;20(4):1029-36. doi: 10.1183/09031936.02.00404202. Eur Respir J. 2002. PMID: 12412700 Review.
-
The mask for noninvasive ventilation: principles of design and effects on aerosol delivery.J Aerosol Med. 2007;20 Suppl 1:S85-98; discussion S98-9. doi: 10.1089/jam.2007.0574. J Aerosol Med. 2007. PMID: 17411410 Review.
Cited by
-
Automatic detection of CO2 rebreathing during BiPAP ventilation.Sci Rep. 2024 Aug 17;14(1):19066. doi: 10.1038/s41598-024-63609-4. Sci Rep. 2024. PMID: 39154044 Free PMC article.
-
Ventilators and Ventilatory Modalities.Front Pediatr. 2020 Sep 2;8:500. doi: 10.3389/fped.2020.00500. eCollection 2020. Front Pediatr. 2020. PMID: 32984212 Free PMC article. Review.
-
What Circuits, Masks and Filters Should Be Used in Home Non-Invasive Mechanical Ventilation.J Clin Med. 2023 Apr 4;12(7):2692. doi: 10.3390/jcm12072692. J Clin Med. 2023. PMID: 37048774 Free PMC article. Review.
-
Year in review in Intensive Care Medicine 2010: III. ARDS and ALI, mechanical ventilation, noninvasive ventilation, weaning, endotracheal intubation, lung ultrasound and paediatrics.Intensive Care Med. 2011 Mar;37(3):394-410. doi: 10.1007/s00134-011-2136-1. Epub 2011 Feb 3. Intensive Care Med. 2011. PMID: 21290103 Free PMC article. No abstract available.
-
Effects of different exhalation valves on CO2 rebreathing and ventilator performance during noninvasive ventilation.Front Med (Lausanne). 2025 Jul 23;12:1538280. doi: 10.3389/fmed.2025.1538280. eCollection 2025. Front Med (Lausanne). 2025. PMID: 40771475 Free PMC article.