Potential errors in pulse oximetry. I. Pulse oximeter evaluation
- PMID: 2014898
- DOI: 10.1111/j.1365-2044.1991.tb09410.x
Potential errors in pulse oximetry. I. Pulse oximeter evaluation
Abstract
There is no absolute reference for oxygen saturation, although multiwavelength in vitro oximeters are accepted as the 'gold standard'. Regardless of whether fractional or functional saturation is used by manufacturers to calibrate their oximeters, evaluation against fractional saturation is recommended since this is the clinically relevant variable. The use of standard notation and comparisons based on bias and precision is recommended. The accuracy of pulse oximetry is intrinsically limited by the use of only two wavelengths, and is dependent on the initial calibration population. The empirical algorithms used to convert the signal to its 'readout value' and the quality control of hardware may both be important sources of variability between oximeters. Change in blood temperature may introduce errors in pulse oximeter and in vitro oximeter saturation readings, but these will be clinically insignificant. Changes in blood pH should not decrease pulse oximetry accuracy.
Similar articles
-
Potential errors in pulse oximetry. II. Effects of changes in saturation and signal quality.Anaesthesia. 1991 Mar;46(3):207-12. doi: 10.1111/j.1365-2044.1991.tb09411.x. Anaesthesia. 1991. PMID: 2014899 Review.
-
Meta-analysis of arterial oxygen saturation monitoring by pulse oximetry in adults.Heart Lung. 1998 Nov-Dec;27(6):387-408. doi: 10.1016/s0147-9563(98)90086-3. Heart Lung. 1998. PMID: 9835670
-
The Sounds of Desaturation: A Survey of Commercial Pulse Oximeter Sonifications.Anesth Analg. 2016 May;122(5):1395-403. doi: 10.1213/ANE.0000000000001240. Anesth Analg. 2016. PMID: 27028772
-
Response of 10 pulse oximeters to an in vitro test system.Br J Anaesth. 1992 Apr;68(4):365-9. doi: 10.1093/bja/68.4.365. Br J Anaesth. 1992. PMID: 1642914
-
Maximizing the laboratory setting for testing devices and understanding statistical output in pulse oximetry.Anesth Analg. 2007 Dec;105(6 Suppl):S85-S94. doi: 10.1213/01.ane.0000268495.35207.ab. Anesth Analg. 2007. PMID: 18048904 Review.
Cited by
-
Use of a Kiosk-Model Self-Triage System for COVID-19 Triage.Risk Manag Healthc Policy. 2025 Feb 21;18:579-592. doi: 10.2147/RMHP.S497053. eCollection 2025. Risk Manag Healthc Policy. 2025. PMID: 40008028 Free PMC article.
-
Pulse Oximetry: The Working Principle, Signal Formation, and Applications.Adv Exp Med Biol. 2022;1384:205-218. doi: 10.1007/978-3-031-06413-5_12. Adv Exp Med Biol. 2022. PMID: 36217086
-
Oxygen saturation measurements using novel diffused reflectance with hyperspectral imaging: Towards facile COVID-19 diagnosis.Opt Quantum Electron. 2022;54(5):322. doi: 10.1007/s11082-022-03658-z. Epub 2022 May 8. Opt Quantum Electron. 2022. PMID: 35571992 Free PMC article.
-
Pulse oximetry: a practical review.BMJ. 1995 Aug 5;311(7001):367-70. doi: 10.1136/bmj.311.7001.367. BMJ. 1995. PMID: 7640545 Free PMC article. Review. No abstract available.
-
Arterial blood gas analysis or oxygen saturation in the assessment of acute asthma?Thorax. 1995 Feb;50(2):186-8. doi: 10.1136/thx.50.2.186. Thorax. 1995. PMID: 7701461 Free PMC article.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources