System accuracy evaluation of 27 blood glucose monitoring systems according to DIN EN ISO 15197
- PMID: 20151773
- DOI: 10.1089/dia.2009.0128
System accuracy evaluation of 27 blood glucose monitoring systems according to DIN EN ISO 15197
Abstract
Background: Blood glucose (BG) monitoring systems enable diabetes patients to effectively control and adjust their therapy. BG monitoring systems with a Conformité Européenne (CE) label should meet the standard DIN EN ISO 15197:2003: > or =95% of the BG results shall fall within +/-15 mg/dL of the reference method at BG concentrations <75 mg/dL and within +/-20% at BG concentrations > or =75 mg/dL. We intended to verify if BG monitoring systems with a CE label fulfill these minimum accuracy requirements.
Methods: We evaluated 27 BG monitoring systems from 18 manufacturers for system accuracy according to DIN EN ISO 15197:2003. Twenty-four systems were compared with the glucose oxidase reaction (YSI 2300 glucose analyzer [YSI Life Sciences, Yellow Springs, OH]) and three systems with the hexokinase reaction (Hitachi 917 [Roche Diagnostics GmbH, Mannheim, Germany]). Duplicate measurements of 100 blood samples with a defined distribution of BG concentrations from 20 mg/dL to 600 mg/dL from > or =100 subjects were included in the evaluation.
Results: Sixteen of the 27 BG monitoring systems fulfilled the minimum accuracy requirements of the standard, i.e., > or =95% of their results showed the minimum acceptable accuracy. Overall, the mean percentage of results showing the minimum acceptable accuracy was 95.2 +/- 5.2%, ranging from 80.0% to 100.0%.
Conclusions: More than 40% of the evaluated BG monitoring systems did not fulfill the minimum accuracy requirements of DIN EN ISO 15197:2003. As inaccurate BG monitoring systems bear the risk of false treatment decisions by the diabetes patient and subsequent possible severe health injury, manufacturers should regularly and effectively check the quality of BG meters and BG test strips.
Similar articles
-
Evaluation of 12 blood glucose monitoring systems for self-testing: system accuracy and measurement reproducibility.Diabetes Technol Ther. 2014 Feb;16(2):113-22. doi: 10.1089/dia.2013.0208. Epub 2013 Nov 8. Diabetes Technol Ther. 2014. PMID: 24205978
-
System accuracy evaluation of 43 blood glucose monitoring systems for self-monitoring of blood glucose according to DIN EN ISO 15197.J Diabetes Sci Technol. 2012 Sep 1;6(5):1060-75. doi: 10.1177/193229681200600510. J Diabetes Sci Technol. 2012. PMID: 23063032 Free PMC article.
-
System Accuracy Evaluation of Four Systems for Self-Monitoring of Blood Glucose Following ISO 15197 Using a Glucose Oxidase and a Hexokinase-Based Comparison Method.J Diabetes Sci Technol. 2015 Apr 14;9(5):1041-50. doi: 10.1177/1932296815580161. J Diabetes Sci Technol. 2015. PMID: 25872967 Free PMC article.
-
Quality of glucose measurement with blood glucose meters at the point-of-care: relevance of interfering factors.Diabetes Technol Ther. 2010 Nov;12(11):847-57. doi: 10.1089/dia.2010.0076. Epub 2010 Sep 30. Diabetes Technol Ther. 2010. PMID: 20879962 Review.
-
Analytical Performance Requirements for Systems for Self-Monitoring of Blood Glucose With Focus on System Accuracy: Relevant Differences Among ISO 15197:2003, ISO 15197:2013, and Current FDA Recommendations.J Diabetes Sci Technol. 2015 Jul;9(4):885-94. doi: 10.1177/1932296815580160. Epub 2015 Apr 14. J Diabetes Sci Technol. 2015. PMID: 25872965 Free PMC article. Review.
Cited by
-
Estimates of total analytical error in consumer and hospital glucose meters contributed by hematocrit, maltose, and ascorbate.J Diabetes Sci Technol. 2010 Nov 1;4(6):1479-94. doi: 10.1177/193229681000400624. J Diabetes Sci Technol. 2010. PMID: 21129345 Free PMC article.
-
Diabetic alert dogs: a preliminary survey of current users.Diabetes Care. 2013 Apr;36(4):e47. doi: 10.2337/dc12-1998. Diabetes Care. 2013. PMID: 23520374 Free PMC article. No abstract available.
-
Accuracy in blood glucose measurement: what will a tightening of requirements yield?J Diabetes Sci Technol. 2012 Mar 1;6(2):435-43. doi: 10.1177/193229681200600232. J Diabetes Sci Technol. 2012. PMID: 22538158 Free PMC article.
-
Acute systemic insulin intolerance does not alter the response of the Akt/GSK-3 pathway to environmental hypoxia in human skeletal muscle.Eur J Appl Physiol. 2015 Jun;115(6):1219-31. doi: 10.1007/s00421-015-3103-2. Epub 2015 Jan 11. Eur J Appl Physiol. 2015. PMID: 25577409
-
ISO-based assessment of accuracy and precision of glucose meters in dogs.J Vet Intern Med. 2014 Sep-Oct;28(5):1405-13. doi: 10.1111/jvim.12397. Epub 2014 Jul 2. J Vet Intern Med. 2014. PMID: 24990398 Free PMC article.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical
Research Materials