Reduction in surface contamination with antineoplastic drugs in 22 hospital pharmacies in the US following implementation of a closed-system drug transfer device
- PMID: 20156932
- PMCID: PMC4627487
- DOI: 10.1177/1078155210361431
Reduction in surface contamination with antineoplastic drugs in 22 hospital pharmacies in the US following implementation of a closed-system drug transfer device
Abstract
Purpose: Surface contamination with the antineoplastic drugs cyclophosphamide, ifosfamide, and 5-fluorouracil was compared in 22 US hospital pharmacies following preparation with standard drug preparation techniques or the PhaSeal® closed-system drug transfer device (CSTD).
Methods: Wipe samples were taken from biological safety cabinet (BSC) surfaces, BSC airfoils, floors in front of BSCs, and counters and analyzed for contamination with cyclophosphamide, ifosfamide, and 5-fluorouracil. Contamination was reassessed several months after the implementation of the CSTD. Surface contamination (ng/cm(2)) was compared between the two techniques and evaluated with the Signed Rank Test.
Results: Using the CSTD compared to the standard preparation techniques, a significant reduction in levels of contamination was observed for all drugs (cyclophosphamide: p < 0.0001; ifosfamide: p < 0.001; 5-fluorouracil: p < 0.01). Median values for surface contamination with cyclophosphamide, ifosfamide, and 5-fluorouracil were reduced by 95%, 90%, and 65%, respectively.
Conclusions: Use of the CSTD significantly reduces surface contamination when preparing cyclophosphamide, ifosfamide, and 5-fluorouracil as compared to standard drug preparation techniques.
Similar articles
-
Comparison of surface contamination with cyclophosphamide and fluorouracil using a closed-system drug transfer device versus standard preparation techniques.Am J Health Syst Pharm. 2006 Sep 15;63(18):1736-44. doi: 10.2146/ajhp050258. Am J Health Syst Pharm. 2006. PMID: 16960258
-
Reduction in Surface Contamination With Cyclophosphamide in 30 US Hospital Pharmacies Following Implementation of a Closed-System Drug Transfer Device.Hosp Pharm. 2013 Mar;48(3):204-12. doi: 10.1310/hpj4803-204. Hosp Pharm. 2013. PMID: 24421463 Free PMC article.
-
Impact and appreciation of two methods aiming at reducing hazardous drug environmental contamination: The centralization of the priming of IV tubing in the pharmacy and use of a closed-system transfer device.J Oncol Pharm Pract. 2014 Dec;20(6):426-32. doi: 10.1177/1078155213517127. Epub 2014 Jan 6. J Oncol Pharm Pract. 2014. PMID: 24395542
-
A scoping review of surface wipe sampling for antineoplastic drug contamination in patient care areas.J Occup Environ Hyg. 2025 Jun;22(6):495-514. doi: 10.1080/15459624.2025.2471397. Epub 2025 Apr 30. J Occup Environ Hyg. 2025. PMID: 40305371
-
Evidence of exposure to cytostatic drugs in healthcare staff: a review of recent literature.Farm Hosp. 2016 Nov 1;40(n06):604-621. doi: 10.7399/fh.2016.40.6.9103. Farm Hosp. 2016. PMID: 27894231 Review. English.
Cited by
-
Reproductive health risks associated with occupational exposures to antineoplastic drugs in health care settings: a review of the evidence.J Occup Environ Med. 2014 Sep;56(9):901-10. doi: 10.1097/JOM.0000000000000249. J Occup Environ Med. 2014. PMID: 25153300 Free PMC article. Review.
-
Surface wipe sampling for antineoplastic (chemotherapy) and other hazardous drug residue in healthcare settings: Methodology and recommendations.J Occup Environ Hyg. 2016 Sep;13(9):658-67. doi: 10.1080/15459624.2016.1165912. J Occup Environ Hyg. 2016. PMID: 27019141 Free PMC article.
-
Effectiveness of a Closed-System Transfer Device in Reducing Surface Contamination in a New Antineoplastic Drug-Compounding Unit: A Prospective, Controlled, Parallel Study.PLoS One. 2016 Jul 8;11(7):e0159052. doi: 10.1371/journal.pone.0159052. eCollection 2016. PLoS One. 2016. PMID: 27391697 Free PMC article.
-
A simple approach to assess the cancer risk of occupational exposure to genotoxic drugs in healthcare settings.J Occup Med Toxicol. 2022 Apr 1;17(1):8. doi: 10.1186/s12995-022-00349-z. J Occup Med Toxicol. 2022. PMID: 35365163 Free PMC article.
-
Exposure Assessment and Monitoring of Antiblastic Drugs Preparation in Health Care Settings: A Systematic Review.Med Lav. 2024 Apr 24;115(2):e2024012. doi: 10.23749/mdl.v115i2.15609. Med Lav. 2024. PMID: 38686575 Free PMC article.
References
-
- Sessink PJM, Bos RP. Drugs hazardous to healthcare workers. Drug Safety. 1999;20:347–359. - PubMed
-
- Acampora A, Castiglia L, Miraglia N, et al. A case study: Surface contamination of cyclophosphamide due to working practices and cleaning procedures in two Italian hospitals. Ann Occup Hyg. 2005;49:611–618. - PubMed
-
- Connor TH, Anderson RW, Sessink PJM, et al. Surface contamination with antineoplastic agents in six cancer treatment centers in Canada and the United States. Am J Health-Syst Pharm. 1999;56:1427–1432. - PubMed
-
- Connor TH, Anderson RW, Sessink PJ, et al. Effectiveness of a closed-system device in containing surface contamination with cyclophosphamide and ifosfamide in an i.v. admixture area. Am J Health-Syst Pharm. 2002;59:68–72. - PubMed
-
- Crauste-Manciet S, Sessink PJM, Ferrari S, et al. Environmental contamination with cytotoxic drugs in healthcare using positive air pressure isolators. Ann Occup Hyg. 2005;49:619–628. - PubMed
Publication types